. 1999 Feb;12(1):12-22.
doi: 10.1007/BF03168622.

The effect of the antiscatter grid on full-field digital mammography phantom images

Affiliations

The effect of the antiscatter grid on full-field digital mammography phantom images

D P Chakraborty. J Digit Imaging. 1999 Feb.

Abstract

Computer Analysis of Mammography Phantom Images (CAMPI) is a method for making quantitative measurements of image quality. This article reports on a recent application of this method to a prototype full-field digital mammography (FFDM) machine. Images of a modified ACR phantom were acquired on the General Electric Diagnostic Molybdenum Rhodium (GE-DMR) FFDM machine at a number of x-ray techniques, both with and without the scatter reduction grid. The techniques were chosen so that one had sets of grid and non-grid images with matched doses (200 mrads) and matched gray-scale values (1500). A third set was acquired at constant 26 kVp and varying mAs for both grid conditions. Analyses of the images yielded signal-to-noise-ratio (SNR), contrast and noise corresponding to each target object, and a non-uniformity measure. The results showed that under conditions of equal gray-scale value the grid images were markedly superior, albeit at higher doses than the non-grid images. Under constant dose conditions, the non-grid images were slightly superior in SNR (7%) but markedly less uniform (60%). Overall, the grid images had substantially greater contrast and superior image uniformity. These conclusions applied to the whole kVp range studied for the Mo-Mo target filter combination and 4 cm of breast equivalent material of average composition. These results suggest that use of the non-grid technique in digital mammography with the GE-DMR-FFDM unit, is presently not warranted. With improved uniformity correction procedure, this conclusion would change and one should be able to realize a 14% reduction in patient dose at the same SNR by using a non-grid technique.

PubMed Disclaimer

References

    1. Feig SA, Yaffe MJ. Current Status of Digital Mammography. Semin Ultrasound, CT and MR. 1996;17:424–43. doi: 10.1016/S0887-2171(96)90029-3. - DOI - PubMed
    1. Eckert MP, Chakraborty DP. Quantitative Analysis of Phantom Images in Mammography. Proc SPIE. 1994;2167:887–899. doi: 10.1117/12.175127. - DOI
    1. Chakraborty DP, Eckert MP. Quantitative versus Subjective Evaluation of Mammography Phantom Images. Med Phys. 1995;22:133–143. doi: 10.1118/1.597463. - DOI - PubMed
    1. Chakraborty DP. Physical measures of image quality in mammography. Proc SPIE. 1996;2708:179–193. doi: 10.1117/12.237781. - DOI
    1. Chakraborty DP. Computer analysis of mammography phantom images (CAMPI) Proc SPIE. 1997;3032:292–299. doi: 10.1117/12.273996. - DOI - PubMed

Publication types