Long-term complications of silicone and hydrogel explants in retinal reattachment surgery
- PMID: 10037564
- DOI: 10.1001/archopht.117.2.197
Long-term complications of silicone and hydrogel explants in retinal reattachment surgery
Abstract
Objective: To compare long-term complications of silicone sponge, silicone rubber, and MIRAgel used as episcleral buckling elements.
Methods: Medical reports were reviewed of 805 patients with cryotherapy and episcleral buckle for rhegmatogenous retinal detachment who were operated on by 1 of us (M.R.-P.) between March 1984 and December 1997. Average follow-up was 76 months. Symptoms and signs of infection or rejection were considered. Care was taken in buckling element removal, considering the material used for scleral buckling (detailed operative note), duration of the buckle, cause of removal, and culture of the removed element.
Results: A total of 757 patients were included in the study. Removal of the implant was necessary in 10 patients (1.3%). Silicone sponge (3 [9%] of 32 patients) was more frequently removed than was silicone rubber (2 [0.6%] of 360 patients) or MIRAgel (5 [1.3%] of 386 patients). Silicone sponge needed to be removed a short time after surgery, showing symptoms of acute infection and positive cultures. Silicone rubber was removed 1 year after surgery with symptoms of chronic infection and positive cultures, and MIRAgel implants were removed after long-term follow-up (7-10 years), showing positive cultures in only 20%.
Conclusion: Periodic long-term follow-up previously recommended for use of other materials also must be recommended for MIRAgel use because of long-term alterations in its chemical composition and eventual swelling of material.
Comment in
-
Histological evidence of hydrogel fragmentation.Arch Ophthalmol. 1999 Oct;117(10):1449. Arch Ophthalmol. 1999. PMID: 10532471 No abstract available.
Similar articles
-
Late swelling and removal of Miragel buckles: a comparison with silicone indentations.Retina. 2003 Oct;23(5):641-6. doi: 10.1097/00006982-200310000-00006. Retina. 2003. PMID: 14574248
-
An unusual complication of retinal reattachment surgery.Ophthalmic Surg Lasers. 1999 Jun;30(6):483-4. Ophthalmic Surg Lasers. 1999. PMID: 10392738
-
Scleral and episcleral histological changes related to encircling explants in 20 eyes.Acta Ophthalmol Scand. 1999 Jun;77(3):279-85. doi: 10.1034/j.1600-0420.1999.770307.x. Acta Ophthalmol Scand. 1999. PMID: 10406146
-
Intraocular intrusion of a scleral sponge implant.Ophthalmic Surg Lasers Imaging. 2003 May-Jun;34(3):223-5. Ophthalmic Surg Lasers Imaging. 2003. PMID: 12757100
-
Complications Assoicated with MIRAgel for Treatment of Retinal Detachment.Semin Ophthalmol. 2018;33(1):89-94. doi: 10.1080/08820538.2017.1353822. Epub 2017 Nov 16. Semin Ophthalmol. 2018. PMID: 29144829 Review.
Cited by
-
Scleral buckle hemorrhagic cyst masquerading as an orbital tumor.Clin Ophthalmol. 2015 Feb 13;9:343-5. doi: 10.2147/OPTH.S76351. eCollection 2015. Clin Ophthalmol. 2015. PMID: 25709397 Free PMC article.
-
Challenges in Diagnosing Extruded Hydrogel Scleral Buckle Mimicking an Orbital Abscess: A Case Report.Cureus. 2024 Mar 18;16(3):e56371. doi: 10.7759/cureus.56371. eCollection 2024 Mar. Cureus. 2024. PMID: 38633963 Free PMC article.
-
A Case of Extrusion of a Solid Silicone Tire Migrating through the Superior Rectus Muscle with Aeromonas hydrophila Infection following a Scleral Buckling Procedure.Case Rep Ophthalmol Med. 2012;2012:167164. doi: 10.1155/2012/167164. Epub 2012 Nov 20. Case Rep Ophthalmol Med. 2012. PMID: 23213585 Free PMC article.
-
An intractable case of Pseudomonas aeruginosa infection after scleral buckling for rhegmatogenous retinal detachment.Clin Ophthalmol. 2008 Mar;2(1):223-6. doi: 10.2147/opth.s2423. Clin Ophthalmol. 2008. PMID: 19668410 Free PMC article.
-
Transpalpebral extrusion of solid silicone buckle.Oman J Ophthalmol. 2009 May;2(2):89-90. doi: 10.4103/0974-620X.53040. Oman J Ophthalmol. 2009. PMID: 20671837 Free PMC article.