Anaesthetic breathing circuit obstruction due to blockage of tracheal tube connector by a foreign body--two cases
- PMID: 10050230
- DOI: 10.1177/0310057X9902700116
Anaesthetic breathing circuit obstruction due to blockage of tracheal tube connector by a foreign body--two cases
Abstract
Two cases are presented which illustrate the disastrous consequences possible when an anaesthetic breathing circuit is obstructed by a foreign body. Despite reports of previous similar cases, work practices and equipment manufacture or design continue to allow for such events to occur. The importance of both pre-anaesthetic testing of the entire circuit including attachments such as the tracheal tube connector and filters, and the removal of these parts should obstruction occur, is emphasised. Use of "clear" transparent breathing circuit components and opaque or brightly coloured packaging and caps which could potentially cause obstruction should decrease the incidence and facilitate the diagnosis of this problem.
Similar articles
-
Alcohol increases bonding between the tracheal tube and the tracheal tube connector.A A Case Rep. 2015 May 1;4(9):114-6. doi: 10.1213/XAA.0000000000000139. A A Case Rep. 2015. PMID: 25909775
-
Defective tracheal tube connector.Can Anaesth Soc J. 1982 Jul;29(4):404. Can Anaesth Soc J. 1982. PMID: 7104814 No abstract available.
-
Another case of obstruction to an anaesthetic circuit.Br J Anaesth. 2003 Jan;90(1):110-1. Br J Anaesth. 2003. PMID: 12488399 No abstract available.
-
[The laryngeal mask].Rev Esp Anestesiol Reanim. 1992 Jan-Feb;39(1):28-33. Rev Esp Anestesiol Reanim. 1992. PMID: 1598447 Review. Spanish. No abstract available.
-
The laryngeal tube.Br J Anaesth. 2005 Dec;95(6):729-36. doi: 10.1093/bja/aei269. Br J Anaesth. 2005. PMID: 16286348 Review.
Cited by
-
Tight reservoir bag: the bag itself may be the culprit.J Clin Monit Comput. 2010 Jun;24(3):189-90. doi: 10.1007/s10877-010-9232-6. Epub 2010 Apr 23. J Clin Monit Comput. 2010. PMID: 20414710
-
Breathing circuit obstruction caused by kink in the reinforced kink-resistant circle system tube.Indian J Anaesth. 2013 Jan;57(1):96-7. doi: 10.4103/0019-5049.108593. Indian J Anaesth. 2013. PMID: 23716787 Free PMC article. No abstract available.
-
Failure to ventilate: Technical error.Saudi J Anaesth. 2013 Jan;7(1):99-100. doi: 10.4103/1658-354X.109853. Saudi J Anaesth. 2013. PMID: 23717246 Free PMC article. No abstract available.
-
A curious case of blocked catheter mount.Med J Armed Forces India. 2013 Jul;69(3):313-4. doi: 10.1016/j.mjafi.2013.03.012. Med J Armed Forces India. 2013. PMID: 24600133 Free PMC article. No abstract available.
-
Manufacturing defect in an endotracheal tube connector: risk of foreign body aspiration.J Anesth. 2008;22(3):333-4. doi: 10.1007/s00540-008-0628-0. Epub 2008 Aug 7. J Anesth. 2008. PMID: 18685950 No abstract available.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Miscellaneous