Considerations in the determination of boron at low concentrations
- PMID: 10050904
- DOI: 10.1007/BF02783122
Considerations in the determination of boron at low concentrations
Abstract
Experimental evidence now supports the nutritional essentiality of boron (B) in some biological systems, and accordingly, the need for reliable analytical B data is increasing. However, the accurate determination of B in biological materials is a formidable challenge at low concentrations (<1 mg B/kg). Recent studies still show significant analytical discrepancies in the analysis of animal tissues and fluids, despite the development of instrumental techniques such as TIMS, ICP-MS, ICP-ES, ICAP, SIMS, NA-MS, PGAA, NRA, and so forth, which have demonstrated detection limits approaching or exceeding (microg B/kg concentrations. Since boric acid is both volatile and ubiquitous in nature, the chemical and physical pathways for B contamination and its loss are manifold, especially during sample preparation. An added obstacle is the inadequacy of biological reference materials certified for B below mg B/kg. With an emphasis toward sample preparation and ICP-MS analysis, examples are provided in this article to help the analyst avoid common problems associated with the analysis of B from biological sources. Topics that are discussed include contamination from Teflon vessels during microwave digestion, losses owing to freeze-drying, B isotopic variations, standards preparation, reagent backgrounds, and instrumental interferences.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources