Development of respiratory gated myocardial SPECT system
- PMID: 10070837
- DOI: 10.1016/s1071-3581(99)90061-2
Development of respiratory gated myocardial SPECT system
Abstract
Background: The superposition of the diaphragm and abdominal structures on the inferior wall of the left ventricle has often distorted single photon emission computed tomography (SPECT). We developed a respiratory gated SPECT (RGS) system to diminish artifacts caused by overlap between the inferior wall and upper abdomen and have validated its feasibility for clinical use.
Methods and results: A 2-detector SPECT system equipped with a respiratory monitor based on impedance plethysmography and an original triggering apparatus was used for RGS in 7 healthy male volunteers. A pulse triggered 100 ms after every expiratory peak was processed in a SPECT system as well as an electrocardiogram (ECG) gating pulse. Inspiratory and expiratory frames were determined using the respiratory curve derived from fluctuation of the gall bladder uptake. Both sets of images were reoriented into short-axis and vertical long-axis slices. For quantification, data were reconstructed into polar plots and count density estimated in 9 myocardial segments. The mean percentage uptake of inferior segments at inspiration was significantly greater than that at expiration (81+/-8.3 versus 76+/-7.1; P < .0001). The inferior-lateral activity ratio improved from 0.78 at expiration to 0.81 at inspiration (P < .01). The coefficient of variance for each segment of inspiratory data was significantly smaller than that at expiration, indicating improved homogeneity of tracer distribution. The lowest cutoff threshold of the tomograms to separate the inferior uptake from that of the upper abdomen was significantly lower at inspiration than at expiration, suggesting smaller scatter from abdominal structures on inspiratory images.
Conclusions: RGS yielded improved tracer uptake of the inferior wall in healthy male subjects and may be suitable as an alternative method for attenuation and scatter correction. However, further clinical validation is needed.
Similar articles
-
Simultaneous transmission-emission thallium-201 cardiac SPECT: effect of attenuation correction on myocardial tracer distribution.J Nucl Med. 1995 Jun;36(6):921-31. J Nucl Med. 1995. PMID: 7769447
-
Effect of attenuation correction on myocardial thallium-201 distribution in patients with a low likelihood of coronary artery disease.Eur J Nucl Med. 1997 Mar;24(3):266-75. doi: 10.1007/BF01728762. Eur J Nucl Med. 1997. PMID: 9143463
-
The value of attenuation correction by hybrid SPECT/CT imaging on infarct size quantification in male patients with previous inferior myocardial infarct.Nucl Med Commun. 2011 Nov;32(11):1026-32. doi: 10.1097/MNM.0b013e32834b4575. Nucl Med Commun. 2011. PMID: 21897308
-
Correction of nonuniform attenuation and image fusion in SPECT imaging by means of separate X-ray CT.Ann Nucl Med. 2002 Jun;16(4):255-61. doi: 10.1007/BF03000104. Ann Nucl Med. 2002. PMID: 12126095
-
Study of Attenuation Correction Using a Cardiac Dynamic Phantom: Synchronized Time-Phase-Gated Attenuation Correction Method.J Nucl Med Technol. 2024 Jun 5;52(2):121-131. doi: 10.2967/jnmt.123.266785. J Nucl Med Technol. 2024. PMID: 38627013
Cited by
-
Have SPECT artifacts breathed their last?J Nucl Cardiol. 2017 Apr;24(2):708-710. doi: 10.1007/s12350-016-0428-0. Epub 2016 Feb 18. J Nucl Cardiol. 2017. PMID: 26892252 No abstract available.
-
A Simulation Study of the Effect of Phase-Shift on Dual Gated Myocardial Perfusion ECT.IEEE Nucl Sci Symp Conf Rec (1997). 2011 Oct;2011:2728-2732. doi: 10.1109/NSSMIC.2011.6152957. IEEE Nucl Sci Symp Conf Rec (1997). 2011. PMID: 26536654 Free PMC article.
-
Respiratory motion detection and correction in ECG-gated SPECT: a new approach.Korean J Radiol. 2008 Nov-Dec;9(6):490-7. doi: 10.3348/kjr.2008.9.6.490. Korean J Radiol. 2008. PMID: 19039264 Free PMC article.
-
Data-driven respiratory motion tracking and compensation in CZT cameras: a comprehensive analysis of phantom and human images.J Nucl Cardiol. 2015 Apr;22(2):308-18. doi: 10.1007/s12350-014-9963-8. Epub 2014 Aug 14. J Nucl Cardiol. 2015. PMID: 25120132
-
Effect of respiratory motion on cardiac defect contrast in myocardial perfusion SPECT: a physical phantom study.Ann Nucl Med. 2019 May;33(5):305-316. doi: 10.1007/s12149-019-01335-y. Epub 2019 Jan 24. Ann Nucl Med. 2019. PMID: 30680536 Free PMC article.
References
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources