. 1999 Jan;170(1):54-60.
doi: 10.1055/s-2007-1011007.

[Accuracy of right and left ventricular heart volume and left ventricular muscle mass determination with cine MRI in breath holding technique]

[Article in German]
Affiliations

[Accuracy of right and left ventricular heart volume and left ventricular muscle mass determination with cine MRI in breath holding technique]

[Article in German]
M B Rominger et al. Rofo. 1999 Jan.

Abstract

Purpose: To assess the accuracy of right- (RV) and left-ventricular (LV) heart volume and muscle mass determinations by cine MR imaging.

Materials and methods: Ten subjects were examined twice by a segmented, ECG-triggered cine sequence in the short axis plane and breath-hold technique by two independent operators and again 4 weeks later by one of the operators. The contours were evaluated manually by two independent examiners and again four weeks later by one of the examiners. LV and RV end-diastolic and end-systolic volumes, stroke volume, cardiac output, ejection fraction (EF), and left-ventricular muscle mass were compared.

Results: The RV evaluation variations (RV-EF: 6.46%) were higher than the LV (LV-EF: 4.46%). The intra-examiner variations were smaller than the inter-examiner variations. The generally operators did not cause a significantly increased variation (LV-EF: 5.77%). In contrast, the repeat examinations at different times did lead to a significant increase in the variation (LV-EF: 10.15%).

Conclusions: The rapid ECG-triggered cine MRI in breath-hold technique allows the simultaneous RV and LV, artefact-free determination of heart volumes and LV muscle mass. The LV volumes can be determined more accurately than the RV. The examiner has a distinct influence on the accuracy, the operator has a lesser effect. Physiological fluctuations of cardiac activity significantly influence the follow-up.