The spectrum and intensity of extra-focal (off-focus) radiation
- PMID: 1009311
- DOI: 10.1259/0007-1285-49-587-951
The spectrum and intensity of extra-focal (off-focus) radiation
Abstract
Conflicting results have been obtained by other workers for the quality of radiation originating outside the area of the focal spot on X-ray tubes. This extra-focal radiation results in poorer picture quality, especially in cases where grids are not being used. As part of an investigation to assess the importance in diagnostic radiology of the various factors affecting the output from X-ray tubes, the photon spectrum has been examined using a Ge(Li) detector. The measured spectra have been adjusted for the effect of the detector characteristics. Results show that the extra-focal radiation is softer than that from the focal spot and, for the X-ray generator measured, constitutes approximately 8% of the exposure at 100 kV with a total filtration equivalent to 2 mm of aluminium. The possibility of further collimation to reduce this component, thus improving picture quality, is considered.
Similar articles
-
Quantifying the effect of off-focal radiation on the output of kilovoltage x-ray systems.Med Phys. 2008 Sep;35(9):4149-60. doi: 10.1118/1.2966348. Med Phys. 2008. PMID: 18841868
-
Penetrameter cassette calibration to 400 kV and effects of extra-focal radiation when measuring tube filtration.Br J Radiol. 1978 Jan;51(601):29-34. doi: 10.1259/0007-1285-51-601-29. Br J Radiol. 1978. PMID: 620116
-
An examination of automatic exposure control regimes for two digital radiography systems.Phys Med Biol. 2009 Aug 7;54(15):4645-70. doi: 10.1088/0031-9155/54/15/002. Epub 2009 Jul 10. Phys Med Biol. 2009. PMID: 19590115
-
(HP)Ge Measurement of spectra for diagnostic X-ray beams.Radiat Prot Dosimetry. 2004;111(1):105-10. doi: 10.1093/rpd/nch370. Radiat Prot Dosimetry. 2004. PMID: 15367779
-
Clinical applications of basic x-ray physics principles.Radiographics. 1998 May-Jun;18(3):731-44; quiz 729. doi: 10.1148/radiographics.18.3.9599394. Radiographics. 1998. PMID: 9599394 Review.
Cited by
-
Modelling the penumbra in Computed Tomography1.J Xray Sci Technol. 2016 May 21;24(4):583-97. doi: 10.3233/XST-160576. J Xray Sci Technol. 2016. PMID: 27232198 Free PMC article.
-
Nondestructive measurement of the grid ratio using a single image.Med Phys. 2009 Oct;36(10):4720-5. doi: 10.1118/1.3191245. Med Phys. 2009. PMID: 19928103 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources
Research Materials