Artefactual elevation of an automated white cell count following femoral vein puncture
- PMID: 10197267
- DOI: 10.1046/j.1365-2257.1999.00178.x
Artefactual elevation of an automated white cell count following femoral vein puncture
Abstract
We report the novel occurrence of an erroneous automated total and differential white blood cell count on a blood sample obtained by traumatic femoral venepuncture, caused by contamination of the blood sample with subcutaneous adipose tissue. The erroneous counts were observed on a Bayer-Technicon H2 analyser while counts on a Coulter GenS, were much less affected. Characteristic scatter plots from both instruments are illustrated.
Comment in
-
Artefactual elevation of the automated white cell count in the context of a bone marrow aspirate analysis.Clin Lab Haematol. 2000 Feb;22(1):56-7. doi: 10.1046/j.1365-2257.2000.00258-2.x. Clin Lab Haematol. 2000. PMID: 10819613 No abstract available.