Long-term follow-up of upper extremity implanted venous access devices in oncology patients
- PMID: 10229476
- DOI: 10.1016/s1051-0443(99)70066-7
Long-term follow-up of upper extremity implanted venous access devices in oncology patients
Abstract
Purpose: To evaluate function and assess incidence of complications relating to upper extremity implanted venous access devices placed in oncology patients primarily for chemotherapy.
Materials and methods: The authors retrospectively evaluated the clinical course of 205 upper extremity implanted venous access devices placed with fluoroscopic and sonographic guidance in 204 patients during a 2-year period. All patients had a diagnosis of malignancy for which chemotherapy was planned. Follow-up data were collected by patient examination, direct evaluation of device function, as well as chart review and review of relevant imaging procedures. A modified technique for device placement is described.
Results: The devices were placed successfully on the initial attempt in all cases. Clinical follow-up was obtained for 195 devices (95.1%) for a total device service period of 33,619 days (mean service interval = 169 days). Seventy-eight devices (40%) had service intervals greater than 180 days. Thirty-seven devices (19% of total devices) led to 39 complications (0.116 event/100 days). No immediate procedural complications were incurred. Eight complications occurred after 180 days of port service. Nineteen devices (9.7% of total devices followed) required removal as a result of complication. Common complications included port malfunction requiring urokinase to clear (n = 10; 0.030 event/100 days), ipsilateral upper extremity deep venous thrombosis (n = 9; 0.027), and local infection (n = 7, 0.021). A comparison of these results relative to other published series of similar devices placed for mixed indications is presented.
Conclusions: Implanted venous access devices are an effective means of long-term venous access in oncology patients. Complication rates in this large series compared favorably to other published radiologic and surgical series. Analysis of complications in a subgroup of extended use implanted venous access devices (greater than 180 days follow-up) showed no statistically significant (P < .05) difference from the larger group of devices.
Comment in
-
Re: long-term follow-up of upper extremity implanted venous access devices in oncology patients.J Vasc Interv Radiol. 1999 Oct;10(9):1281. doi: 10.1016/s1051-0443(99)70232-0. J Vasc Interv Radiol. 1999. PMID: 10527209 No abstract available.
Similar articles
-
Incidence and management of catheter occlusion in implantable arm ports: results in 391 patients.J Vasc Interv Radiol. 1999 Jun;10(6):767-74. doi: 10.1016/s1051-0443(99)70112-0. J Vasc Interv Radiol. 1999. PMID: 10392945
-
Infectious complications of 393 peripherally implantable venous access devices in HIV-positive and HIV-negative patients.J Vasc Interv Radiol. 1999 Jan;10(1):71-7. doi: 10.1016/s1051-0443(99)70014-x. J Vasc Interv Radiol. 1999. PMID: 10872493
-
Long-term, totally implantable central venous access ports connected to a Groshong catheter for chemotherapy of solid tumours: experience from 178 cases using a single type of device.Eur J Cancer. 1997 Jul;33(8):1190-4. doi: 10.1016/s0959-8049(97)00039-7. Eur J Cancer. 1997. PMID: 9301441 Clinical Trial.
-
[Fibrinolysis of deep venous thrombosis on implantable perfusion devices. Apropos of a consecutive series of 57 cases of thrombosis and 32 cases of fibrinolysis].Bull Cancer. 1996 Apr;83(4):293-9. Bull Cancer. 1996. PMID: 8680080 Review. French.
-
Central venous access port devices - a pictorial review of common complications from the interventional radiology perspective.J Vasc Access. 2012 Jan-Mar;13(1):9-15. doi: 10.5301/JVA.2011.8439. J Vasc Access. 2012. PMID: 21725953 Review.
Cited by
-
Frequency of pneumothorax and haemothorax after primary open versus closed implantation strategies for insertion of a totally implantable venous access port in oncological patients: study protocol for a randomised controlled trial.Trials. 2015 Mar 31;16:128. doi: 10.1186/s13063-015-0643-z. Trials. 2015. PMID: 25872780 Free PMC article. Clinical Trial.
-
Retrospective analysis: 5509 cases of "totally implantable venous access port systems implantation (TIVAPS) depth" assisted by digital radiography.Langenbecks Arch Surg. 2022 Nov;407(7):3123-3132. doi: 10.1007/s00423-022-02573-x. Epub 2022 Jun 3. Langenbecks Arch Surg. 2022. PMID: 35660962
-
Central venous access: techniques and indications in oncology.Eur Radiol. 2008 Oct;18(10):2333-44. doi: 10.1007/s00330-008-0981-2. Epub 2008 May 6. Eur Radiol. 2008. PMID: 18458909 Review.
-
Upper arm central venous port implantation: a 6-year single institutional retrospective analysis and pictorial essay of procedures for insertion.PLoS One. 2014 Mar 10;9(3):e91335. doi: 10.1371/journal.pone.0091335. eCollection 2014. PLoS One. 2014. PMID: 24614412 Free PMC article.
-
Outcome analysis in 3,160 implantations of radiologically guided placements of totally implantable central venous port systems.Eur Radiol. 2011 Jun;21(6):1224-32. doi: 10.1007/s00330-010-2045-7. Epub 2011 Jan 5. Eur Radiol. 2011. PMID: 21207035
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources