Pacemaker lead extraction with the laser sheath: results of the pacing lead extraction with the excimer sheath (PLEXES) trial
- PMID: 10334441
- DOI: 10.1016/s0735-1097(99)00074-1
Pacemaker lead extraction with the laser sheath: results of the pacing lead extraction with the excimer sheath (PLEXES) trial
Abstract
Objectives: The purpose of this study was to evaluate the safety and effectiveness of pacemaker lead extraction with the excimer sheath in comparison to nonlaser lead extraction.
Background: Fibrotic attachments that develop between chronically implanted pacemaker leads and to the venous, valvular and cardiac structures are the major obstacles to safe and consistent lead extraction. Locking stylets and telescoping sheaths produce a technically demanding but effective technique of mechanically disrupting the fibrosis. However, ultraviolet excimer laser light dissolves instead of tearing the tissue attachments.
Methods: A randomized trial of lead extraction was conducted in 301 patients with 465 chronically implanted pacemaker leads. The laser group patients had the leads removed with identical tools as the nonlaser group with the exception that the inner telescoping sheath was replaced with the 12-F excimer laser sheath. Success for both groups was defined as complete lead removal with the randomized therapy without complications.
Results: Complete lead removal rate was 94% in the laser group and 64% in the nonlaser group (p = 0.001). Failed nonlaser extraction was completed with the laser tools 88% of the time. The mean time to achieve a successful lead extraction was significantly reduced for patients randomized to the laser tools, 10.1 +/- 11.5 min compared with 12.9 +/- 19.2 min for patients randomized to nonlaser techniques (p < 0.04). Potentially life-threatening complications occurred in none of the nonlaser and three of the laser patients, including one death (p = NS).
Conclusions: Laser-assisted pacemaker lead extraction has significant clinical advantages over extraction without laser tools and is associated with significant risks.
Similar articles
-
Initial experience in the extraction of chronically implanted pacemaker leads using the Excimer laser sheath.Heart. 1999 Jul;82(1):101-4. doi: 10.1136/hrt.82.1.101. Heart. 1999. PMID: 10377320 Free PMC article.
-
First European experience using excimer laser for the extraction of permanent pacemaker leads.Pacing Clin Electrophysiol. 1998 Jan;21(1 Pt 2):268-70. doi: 10.1111/j.1540-8159.1998.tb01102.x. Pacing Clin Electrophysiol. 1998. PMID: 9474686
-
Excimer laser for pacemaker and defibrillator lead extraction: techniques and clinical results.Lasers Med Sci. 2001;16(2):113-21. doi: 10.1007/pl00011340. Lasers Med Sci. 2001. PMID: 11484752
-
Intravascular lead extraction using the excimer laser: pitfalls and tips for success.Semin Vasc Surg. 2008 Mar;21(1):54-6. doi: 10.1053/j.semvascsurg.2007.11.004. Semin Vasc Surg. 2008. PMID: 18342737 Review.
-
Active fixation mechanism complicates coronary sinus lead extraction and limits subsequent reimplantation targets.J Interv Card Electrophysiol. 2013 Jan;36(1):81-6; discussion 86. doi: 10.1007/s10840-012-9704-3. Epub 2012 Aug 7. J Interv Card Electrophysiol. 2013. PMID: 22869386 Review.
Cited by
-
Is cardiopulmonary bypass standby still required for laser lead extractions?J Cardiothorac Surg. 2022 Sep 15;17(1):235. doi: 10.1186/s13019-022-01987-4. J Cardiothorac Surg. 2022. PMID: 36109812 Free PMC article.
-
Lead extraction using a laser system: Techniques, efficacy, and limitations.J Arrhythm. 2016 Aug;32(4):279-82. doi: 10.1016/j.joa.2015.06.006. Epub 2015 Aug 18. J Arrhythm. 2016. PMID: 27588149 Free PMC article. Review.
-
Cardiac Implantable Electronic Device Infection in Patients at Risk.Arrhythm Electrophysiol Rev. 2016 May;5(1):65-71. doi: 10.15420/aer.2015.27.2. Arrhythm Electrophysiol Rev. 2016. PMID: 27403296 Free PMC article.
-
Percutaneous treatment of the superior vena cava syndrome via an excimer laser sheath in a patient with a single chamber atrial pacemaker.J Interv Card Electrophysiol. 2006 Sep;16(3):203-6. doi: 10.1007/s10840-006-9041-5. Epub 2006 Dec 13. J Interv Card Electrophysiol. 2006. PMID: 17165133
-
Advances in device lead extraction.Curr Cardiol Rep. 2001 Jul;3(4):324. doi: 10.1007/s11886-001-0087-9. Curr Cardiol Rep. 2001. PMID: 11406091 Review.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical