Accuracy of tympanic temperature readings in children under 6 years of age
- PMID: 10335248
Accuracy of tympanic temperature readings in children under 6 years of age
Abstract
Purpose: Infrared tympanic thermometry (ITT) is increasingly used as a convenient, noninvasive assessment method for febrile children. However, the accuracy of ITT for children has been questioned, particularly in relation to specificity and sensitivity. This study was designed to (a) determine the correlation and extent of agreement between rectal temperature (RT) readings obtained by electronic thermometer and ear-based temperature readings obtained by ITT, and (b) determine the accuracy of detecting fever in children under 6 years of age.
Methods: This correlational study used a sample of 241 paired ear and rectal temperatures obtained in the emergency department (ED) of a 920-bed regional hospital. All children under the age of 6 years who routinely received a rectal temperature measurement were eligible to participate. According to the ED protocol, rectal temperatures were obtained on all patients less than 3 years or patients 3-6 years that presented with a complaint of fever. For the study, tympanic measurements were also taken.
Results: Correlation between rectal and tympanic temperature readings was statistically significant (r = 0.84, p < .001). The mean difference between rectal and tympanic temperatures was -0.60 degrees C. Threshold-adjusted accuracy in screening for fever was determined by sensitivity (80%), specificity (85%), positive predictive value (87%), and negative predictive value (85%).
Conclusions: Sensitivity, specificity, positive predictive value, and negative predictive value are unacceptably low and the number of children with fever who would be missed by screening with a tympanic thermometer is unacceptable. Findings of this study do not support the use of tympanic thermometers to detect fever in children under 6 years of age.
Comment in
-
Questions regarding conclusions in article on tympanic temperature readings.Pediatr Nurs. 1999 Nov-Dec;25(6):619-20; author reply 620. Pediatr Nurs. 1999. PMID: 12024380 No abstract available.
Similar articles
-
[Measurement of fever in children--is infrared tympanic thermometry reliable?].Tidsskr Nor Laegeforen. 2002 Nov 20;122(28):2700-1. Tidsskr Nor Laegeforen. 2002. PMID: 12523088 Norwegian.
-
[Can ear temperature measurement be used in a hospital?].Tidsskr Nor Laegeforen. 2005 Oct 20;125(20):2763-5. Tidsskr Nor Laegeforen. 2005. PMID: 16244675 Norwegian.
-
Accuracy of infrared ear thermometry and traditional temperature methods in young children.Heart Lung. 1994 May-Jun;23(3):181-95. Heart Lung. 1994. PMID: 8039988 Clinical Trial.
-
[Body temperature measurement in daily practice].Arch Pediatr. 2005 Aug;12(8):1292-300. doi: 10.1016/j.arcped.2005.01.034. Arch Pediatr. 2005. PMID: 15993044 Review. French.
-
Is tympanic membrane thermometry the best method for recording temperature in children?J Child Health Care. 2006 Jun;10(2):96-110. doi: 10.1177/1367493506062550. J Child Health Care. 2006. PMID: 16707539 Review.
Cited by
-
Thermometry in children.J Emerg Trauma Shock. 2012 Jul;5(3):246-9. doi: 10.4103/0974-2700.99699. J Emerg Trauma Shock. 2012. PMID: 22988404 Free PMC article.
-
Fever detection in under 5 children in a tertiary health facility using the infrared tympanic thermometer in the oral mode.Ital J Pediatr. 2011 Jan 22;37:8. doi: 10.1186/1824-7288-37-8. Ital J Pediatr. 2011. PMID: 21255450 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Medical