An aiming device for pin fixation at the iliac crest for external fixation in unstable pelvic fracture
- PMID: 10341899
- DOI: 10.1016/s0020-1383(98)00180-6
An aiming device for pin fixation at the iliac crest for external fixation in unstable pelvic fracture
Abstract
To improve the accuracy of pinning at the iliac crest during external fixation of the pelvic fracture, an aiming device has been designed. The device consists of 3 parts: a sleeve which accommodates a 5.0 Shanz pin, a handle and guide points. The guide points were designed to grasp the iliac crest to allow proper pin fixation. The device has undergone trials to fix Shanz pins on the iliac crests of 10 cadavers by 10 recently graduated doctors. All pins were fixed in proper position and passed into the bone between the two tables of the iliac crest without penetrating the tables. The device has so far been used in 50 patients who had unstable pelvic fractures. All pins were in the proper positions and there had been no loosening at the time the pins were removed. The use of this aiming device for pinning the iliac crest for external fixation of pelvic fracture has given encouraging results.
Similar articles
-
Pin placement in pelvic external fixation.Injury. 1993 Oct;24(9):581-4. doi: 10.1016/0020-1383(93)90117-o. Injury. 1993. PMID: 8288374 Clinical Trial.
-
Strength of the pin-bone interface of external fixation pins in the iliac crest. A biomechanical study.Clin Orthop Relat Res. 1995 Jan;(310):237-44. Clin Orthop Relat Res. 1995. PMID: 7641445
-
Effect of pin location on stability of pelvic external fixation.Clin Orthop Relat Res. 1999 Apr;(361):237-44. doi: 10.1097/00003086-199904000-00030. Clin Orthop Relat Res. 1999. PMID: 10212618 Clinical Trial.
-
External fixation of unstable pelvic fractures: a systematic review and meta-analysis.ANZ J Surg. 2019 Sep;89(9):1022-1027. doi: 10.1111/ans.15027. Epub 2019 Feb 12. ANZ J Surg. 2019. PMID: 30756458
-
The Use of External Fixation for the Management of the Unstable Anterior Pelvic Ring.J Orthop Trauma. 2018 Sep;32 Suppl 6:S14-S17. doi: 10.1097/BOT.0000000000001251. J Orthop Trauma. 2018. PMID: 30095676 Review.
Cited by
-
Optimal pin position in supraacetabular external fixation.ANZ J Surg. 2021 Apr;91(4):680-684. doi: 10.1111/ans.16694. Epub 2021 Mar 18. ANZ J Surg. 2021. PMID: 33734540 Free PMC article.
-
Achievable pin spanning angulation in anterosuperior pelvic external fixation.Eur J Orthop Surg Traumatol. 2020 Oct;30(7):1163-1170. doi: 10.1007/s00590-020-02684-5. Epub 2020 May 1. Eur J Orthop Surg Traumatol. 2020. PMID: 32358714
MeSH terms
LinkOut - more resources
Full Text Sources