Atrial sensing and AV synchrony in single lead VDD pacemakers: a prospective comparison to DDD devices with bipolar atrial leads
- PMID: 10355692
- DOI: 10.1111/j.1540-8167.1999.tb00707.x
Atrial sensing and AV synchrony in single lead VDD pacemakers: a prospective comparison to DDD devices with bipolar atrial leads
Abstract
Introduction: Single lead VDD pacing has offered an alternative to DDD systems in patients with isolated AV block. Up to now, however, the relative performance of these pacemaker systems was not systematically compared.
Methods and results: Three hundred sixty patients who received either a VDD pacemaker (n = 180) or a DDD device (n = 180) with a bipolar atrial lead were investigated prospectively for a mean period of 30 +/- 13 months. Pacemaker function was analyzed by telemetry, Holter monitoring, and exercise ECG. Time of implantation and fluoroscopy was significantly lower with VDD devices (44.3 +/- 5.1 min vs 74.4 +/- 13.5 min and 4.6 +/- 2.5 min vs 10.3 +/- 5.6 min in DDD pacemakers, respectively). Intermittent atrial undersensing occurred in 23.3% of patients with a VDD pacemaker and in 9.4% with DDD devices (NS). The incidence of atrial tachyarrhythmias did not differ between the VDD (6.7%) and the DDD group (6.1%). Sinus node dysfunction developed in 1.9% of patients, but the vast majority (85.7%) of patients were asymptomatic. There was a tendency for a higher rate of operative revisions in the DDD group (6.1% vs 3.3% in VDD pacemakers, P = 0.15). Cumulative maintenance of AV-synchronized pacing mode was 94.9% in patients with VDD pacemakers and 92.1% with DDD devices (NS).
Conclusion: With the benefit of a simpler implant procedure, long-term outcome of single lead VDD pacing is equivalent to DDD pacing in patients with AV block and preoperative normal sinus node function.
Similar articles
-
Dual chamber cardiac pacing in children: Single chamber pacing dual chamber sensing cardiac pacemaker or dual chamber pacing and sensing cardiac pacemaker?Pediatr Int. 2002 Dec;44(6):635-40. doi: 10.1046/j.1442-200x.2002.01631.x. Pediatr Int. 2002. PMID: 12421261
-
Efficacy and safety of bipolar sensing with high atrial sensitivity in dual chamber pacemakers.Pacing Clin Electrophysiol. 2000 Apr;23(4 Pt 1):427-33. doi: 10.1111/j.1540-8159.2000.tb00822.x. Pacing Clin Electrophysiol. 2000. PMID: 10793429
-
VDD(R) pacing: short- and long-term stability of atrial sensing with a single lead system.Pacing Clin Electrophysiol. 1996 Apr;19(4 Pt 1):455-64. doi: 10.1111/j.1540-8159.1996.tb06516.x. Pacing Clin Electrophysiol. 1996. PMID: 8848393
-
[Pacemaker therapy in patients with atrial fibrillation].Herz. 1998 Jun;23(4):260-8. doi: 10.1007/BF03044321. Herz. 1998. PMID: 9690113 Review. German.
-
The current status of single lead dual chamber sensing and pacing.J Interv Card Electrophysiol. 1998 Sep;2(3):255-67. doi: 10.1023/a:1009737020626. J Interv Card Electrophysiol. 1998. PMID: 9870020 Review.
Cited by
-
Long term complications in single and dual chamber pacing are influenced by surgical experience and patient morbidity.Heart. 2005 Apr;91(4):500-6. doi: 10.1136/hrt.2003.025411. Heart. 2005. PMID: 15772212 Free PMC article.
-
Häufigkeit atrialen Undersensings bei Patienten mit höhergradigen AV-Blockierungen randomisiert auf VDD-bzw. DDD-Schrittmacher.Herzschrittmacherther Elektrophysiol. 2000 Jan;11 Suppl 1:5-6. doi: 10.1007/BF03042505. Herzschrittmacherther Elektrophysiol. 2000. PMID: 19495621 German. No abstract available.
-
Langzeit-Stabilität von VDD-Elektroden.Herzschrittmacherther Elektrophysiol. 2000 Jan;11 Suppl 1:3-4. doi: 10.1007/BF03042504. Herzschrittmacherther Elektrophysiol. 2000. PMID: 19495620 German. No abstract available.
-
VDD pacemaker replacement is safe and reliable independently of the previously implanted lead: a prospective and controlled study.J Interv Card Electrophysiol. 2006 Mar;15(2):107-11. doi: 10.1007/s10840-006-7743-3. J Interv Card Electrophysiol. 2006. PMID: 16755339 Clinical Trial.
-
Intracardiac atrial signal amplitude in congenital and acquired complete heart block.Indian Pacing Electrophysiol J. 2010 Oct 31;10(10):447-53. Indian Pacing Electrophysiol J. 2010. PMID: 21151383 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical
Research Materials
Miscellaneous