Validity, reliability, and calibration of the Tritrac accelerometer as a measure of physical activity
- PMID: 10378921
- DOI: 10.1097/00005768-199906000-00022
Validity, reliability, and calibration of the Tritrac accelerometer as a measure of physical activity
Abstract
Purpose: The purposes of this study were to assess the validity and reliability of the Tritrac R3D accelerometer during treadmill walking and running and then to calibrate the instrument.
Methods: The Tritrac was assessed on 60 young adults (23.4 +/- 2.9 yr) during treadmill walking and running at 3.2, 6.4, and 9.7 km x h(-1). The calibration was carried out by identifying ranges of Tritrac raw data (vector magnitude) values corresponding to light (2-3.9 MET), moderate (4-7 MET), and vigorous (>7 MET) physical activity. Energy expenditure (EE), measured by indirect calorimetry, served as the criterion measure.
Results: Interinstrument intraclass reliability coefficients for Tritracs worn on the right and left hip ranged from 0.73-0.87, while intersession coefficients demonstrated high reliability for all speeds (R = 0.87-0.92). Paired t-tests comparing mean accelerometer counts at 6.4 km x h(-1), 0% grade (2647 +/- 456), and 6.4 km x h(-1), 5% grade (2635 +/- 435) demonstrated no significant difference (P > 0.05). Mean differences between EE measured by indirect calorimetry and that estimated by the Tritrac ranged from 0.0082 kcal x kg(-1) x min(-1) at 3.2 km x h(-1) to 0.0320 kcal x kg(-1) x min(-1) at 9.7 km x h(-1), with the Tritrac consistently overestimating EE during horizontal treadmill walking. The relationship between vector magnitude and EE across all speeds was highly linear (R2 = 0.90, SEE = 0.014 kcal x kg(-1) x min(-1)), with little overlap between light, moderate, and vigorous categories. The mean vector magnitudes at 2, 4, and 7 MET were 650, 1772, and 3455, respectively.
Conclusions: These data indicate that the Tritrac is highly reliable from day to day and is sensitive to changes in speed but not grade. Furthermore, the Tritrac accurately distinguishes various intensities of walking and jogging on level ground. With limitations, these cut-points can be used to categorize light, moderate, and vigorous physical activity and to estimate EE.
Comment in
-
Energy expenditure study makes erroneous inferences and recommendation.Med Sci Sports Exerc. 2001 Jun;33(6):1060-1. doi: 10.1097/00005768-200106000-00029. Med Sci Sports Exerc. 2001. PMID: 11404675 No abstract available.
Similar articles
-
The accuracy of the TriTrac-R3D accelerometer to estimate energy expenditure.Med Sci Sports Exerc. 1999 May;31(5):747-54. doi: 10.1097/00005768-199905000-00020. Med Sci Sports Exerc. 1999. PMID: 10331898
-
Field evaluation of energy expenditure in women using Tritrac accelerometers.Med Sci Sports Exerc. 2002 Oct;34(10):1667-74. doi: 10.1097/00005768-200210000-00020. Med Sci Sports Exerc. 2002. PMID: 12370570
-
Reliability and validity of the tritrac-R3D accelerometer during backpacking: a case study.Percept Mot Skills. 2001 Aug;93(1):37-46. doi: 10.2466/pms.2001.93.1.37. Percept Mot Skills. 2001. PMID: 11693704
-
Comparison of activity monitors to estimate energy cost of treadmill exercise.Med Sci Sports Exerc. 2004 Jul;36(7):1244-51. doi: 10.1249/01.mss.0000132379.09364.f8. Med Sci Sports Exerc. 2004. PMID: 15235333
-
Calibration of accelerometer output for adults.Med Sci Sports Exerc. 2005 Nov;37(11 Suppl):S512-22. doi: 10.1249/01.mss.0000185659.11982.3d. Med Sci Sports Exerc. 2005. PMID: 16294114 Review.
Cited by
-
Predicting energy expenditure of physical activity using hip- and wrist-worn accelerometers.Diabetes Technol Ther. 2003;5(6):1023-33. doi: 10.1089/152091503322641088. Diabetes Technol Ther. 2003. PMID: 14709206 Free PMC article.
-
Optimal sensor placement for measuring physical activity with a 3D accelerometer.Sensors (Basel). 2014 Feb 18;14(2):3188-206. doi: 10.3390/s140203188. Sensors (Basel). 2014. PMID: 24553085 Free PMC article.
-
Reliability and validity of the Mywellness Key physical activity monitor.Clin Epidemiol. 2013;5:13-20. doi: 10.2147/CLEP.S38370. Epub 2013 Jan 25. Clin Epidemiol. 2013. PMID: 23378783 Free PMC article.
-
Effects of an 8-month exercise training program on off-exercise physical activity.Med Sci Sports Exerc. 2011 Sep;43(9):1744-51. doi: 10.1249/MSS.0b013e3182148a7e. Med Sci Sports Exerc. 2011. PMID: 21364488 Free PMC article. Clinical Trial.
-
Effect of a moderate-intensity demonstration walk on accuracy of physical activity self-report.J Exerc Sci Fit. 2017 Jun;15(1):1-7. doi: 10.1016/j.jesf.2016.10.002. Epub 2016 Dec 18. J Exerc Sci Fit. 2017. PMID: 29541123 Free PMC article.
Publication types
MeSH terms
Grants and funding
LinkOut - more resources
Full Text Sources
Medical
Miscellaneous