Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Comparative Study
. 1999 Jul;45(7):1077-81.

The use of infrared spectrophotometry for measuring body water spaces

Affiliations
  • PMID: 10388485
Comparative Study

The use of infrared spectrophotometry for measuring body water spaces

G Jennings et al. Clin Chem. 1999 Jul.

Abstract

Background: The conventional method of measuring total body water by the deuterium isotope dilution method uses gas isotope ratio mass spectrometry (IRMS), which is both expensive and time-consuming. We investigated an alternative method, using Fourier transform infrared spectrophotometry (FTIR), which uses less expensive instrumentation and requires little sample preparation.

Method: Total body water measurements in human subjects were made by obtaining plasma, saliva, and urine samples before and after oral dosing with 1.5 mol of deuterium oxide. The enrichments of the body fluids were determined from the FTIR spectra in the range 1800-2800 cm-1, using a novel algorithm for estimation of instrumental response, and by IRMS for comparison.

Results: The CV (n = 5) for repeat determinations of deuterium oxide in biological fluids and calibrator solutions (400-1000 micromol/mol) was found to be in the range 0.1-0.9%. The use of the novel algorithm instead of the integration routines supplied with the instrument gave at least a threefold increase in precision, and there was no significant difference between the results obtained with FTIR and those obtained with IRMS.

Conclusion: This improved infrared method for measuring deuterium enrichment in plasma and saliva requires no sample preparation, is rapid, and has potential value to the clinician.

PubMed Disclaimer

Publication types

Substances

LinkOut - more resources