Entry force and intra-abdominal pressure associated with six laparoscopic trocar-cannula systems: a randomized comparison

Affiliations
Clinical Trial

Entry force and intra-abdominal pressure associated with six laparoscopic trocar-cannula systems: a randomized comparison

C M Tarnay et al. Obstet Gynecol. 1999 Jul.

Abstract

Objective: In trocar-cannula systems, increased entry force could result in loss of operator control, a potential cause of serious visceral and vascular injuries. We developed a system to measure entry force and intraperitoneal pressure to evaluate and compare trocar-cannula systems.

Methods: Six laparoscopic trocar-cannula systems of similar diameter (12 mm) were tested (two pyramidal, two cutting-dilating, and two blunt conical) using a white swine model. All six systems were inserted into each of 12 subjects with location designated by random allotment (72 insertions). During each insertion, intraperitoneal pressure and entry force were measured using a system consisting of a gas-gas transducer, a 50-lb load cell, and a multichannel data acquisition board. Mean entry force and intraperitoneal pressure were compared using mixed-model analysis of variance.

Results: Mean entry force measurements were as follows: pyramidal 9.01 lb and 13.48 lb, cutting-dilating 9.94 lb and 16.46 lb, and blunt conical 19.15 lb and 31.91 lb. Intraperitoneal pressure changes generally reflected measured entry force.

Conclusion: The system successfully measured both entry force and resultant intraperitoneal pressure. Pyramidal trocar-cannula systems required the lowest force for entry. These differences in entry force have potential clinical implications related to the risk of visceral and vascular injury. Intraperitoneal pressure measurement could be used as a surrogate for insertional force measurement.