Contrast mechanisms in MR imaging
- PMID: 10415233
- DOI: 10.1007/s003300050789
Contrast mechanisms in MR imaging
Abstract
This paper is a brief introduction to tissue-specific parameters and the utilization of various MR imaging sequences to display these parameters in order to differentiate normal from pathologic tissue and function. The three dominant tissue-specific parameters discussed are proton density, longitudinal relaxation time T1, and transverse relaxation time T2. For the utilization of gradient-echo sequences, transverse relaxation time T2(*) is introduced, more dependent on the environment or tissue interfaces than on the tissue itself. Another tissue-specific parameter is the concentration of macromolecules and their hydration layers as targeted with magnetization transfer imaging. Still another tissue-specific parameter is the chemical environment. Functional parameters that influence the contrast are diffusion, perfusion, flow, or motion. The sequence-related utilization of these tissue-specific parameters start with magnetization preparation as in spectral suppression of fat signal, relaxation-dependent elimination of fat or cerebrospinal fluid (CSF) signal, simple inversion pulse, magnetization transfer saturation, or diffusion weighting. Possible contrast mechanisms for the tissue-specific parameters are discussed for each of the commonly used sequences, whether of spin-echo type or of gradient-echo type, with or without magnetization preparation, conventional single-echo acquisition, or contemporary multiecho acquisition.