Self-tonometry with the Ocuton S versus Goldmann tonometry
- PMID: 10447645
- DOI: 10.1007/s004170050302
Self-tonometry with the Ocuton S versus Goldmann tonometry
Abstract
Background: The purpose of the study was to evaluate the acceptance and reproducibility of a new self-tonometer (Ocuton S) and to compare it with the Goldmann applanation tonometer.
Methods: The Ocuton S was studied in 20 healthy volunteers and 100 patients. After detailed explanation to each individual and a number of test measurements, we studied the reproducibility and accuracy of the Ocuton S in 20 healthy volunteers with and without topical anaesthesia. Acceptance was graded by means of visual analogue scales. Each volunteer had to measure his/her own IOP in three consecutive measurements. The data were compared with Goldmann tonometry at random times. One hundred patients were introduced to the Ocuton S. Either three consecutive self-measurements of IOP were performed and then compared with three consecutive measurements by Goldmann tonometry, or vice versa. This was done randomly.
Results: Forty-one of 100 patients were unable to perform three consecutive measurements with the Ocuton S after at least 15 min introduction time to the device (non-success). Fifty-nine of 100 patients were able to measure their IOP with the Ocuton S. The mean pressure value with the Ocuton S was 23.2+/-8.2 mmHg, compared with Goldmann 18. 4+/-5.7 mmHg. The difference of 5+/-5.7 mmHg was statistically significant (paired t-test p<0.01). The visual analogue scale scores of healthy volunteers (100 mm = maximal comfort, 0 mm = not acceptable), was 72+/-31 in anaesthetised eyes and 39+/-41 mm without anaesthesia.
Conclusion: The Ocuton S seems to be an acceptable means for the majority of patients of measuring their IOP at home. Refinement of the accuracy of the device seems necessary.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
