Plasticity-induced damage layer is a precursor to wear in radiation-cross-linked UHMWPE acetabular components for total hip replacement. Ultra-high-molecular-weight polyethylene
- PMID: 10475563
- DOI: 10.1016/s0883-5403(99)90086-4
Plasticity-induced damage layer is a precursor to wear in radiation-cross-linked UHMWPE acetabular components for total hip replacement. Ultra-high-molecular-weight polyethylene
Abstract
The mechanism for the improved wear resistance of cross-linked ultra-high-molecular-weight polyethylene (UHMWPE) remains unclear. This study investigated the effect of cross-linking achieved by gamma irradiation in nitrogen on the tribologic, mechanical, and morphologic properties of UHMWPE. The goal of this study was to relate UHMWPE properties to the wear mechanism in acetabular-bearing inserts. Wear simulation of acetabular liners was followed by detailed characterization of the mechanical behavior and crystalline morphology at the articulating surface. The wear rate was determined to be directly related to the ductility, toughness, and strain-hardening behavior of the UHMWPE. The concept of a plasticity-induced damage layer is introduced to explain the near-surface orientation of the crystalline lamellae observed in the wear-tested acetabular liners. Cross-linking reduces abrasive wear of acetabular components by substantially reducing--but not eliminating--the plasticity-induced damage layer that precedes abrasive wear.
Similar articles
-
Wear behaviour of cross-linked polyethylene assessed in vitro under severe conditions.Biomaterials. 2005 Jun;26(16):3259-67. doi: 10.1016/j.biomaterials.2004.07.070. Biomaterials. 2005. PMID: 15603821
-
Larger diameter femoral heads used in conjunction with a highly cross-linked ultra-high molecular weight polyethylene: a new concept.J Arthroplasty. 2001 Dec;16(8 Suppl 1):24-30. doi: 10.1054/arth.2001.28376. J Arthroplasty. 2001. PMID: 11742447
-
Long-term wear failure analysis of uhmwpe acetabular cup in total hip replacement.J Mech Behav Biomed Mater. 2018 Nov;87:1-9. doi: 10.1016/j.jmbbm.2018.07.017. Epub 2018 Jul 11. J Mech Behav Biomed Mater. 2018. PMID: 30031358
-
Implant wear mechanisms--basic approach.Biomed Mater. 2008 Dec;3(4):042001. doi: 10.1088/1748-6041/3/4/042001. Epub 2008 Sep 25. Biomed Mater. 2008. PMID: 18824778 Review.
-
[Research progress of backside wear in acetabular liners].Zhongguo Xiu Fu Chong Jian Wai Ke Za Zhi. 2013 Dec;27(12):1453-6. Zhongguo Xiu Fu Chong Jian Wai Ke Za Zhi. 2013. PMID: 24640364 Review. Chinese.
Cited by
-
Feasibility of using diamond-like carbon films in total joint replacements: a review.J Mater Sci Mater Med. 2024 Aug 13;35(1):47. doi: 10.1007/s10856-024-06814-x. J Mater Sci Mater Med. 2024. PMID: 39136884 Free PMC article. Review.
-
Can pin-on-disk testing be used to assess the wear performance of retrieved UHMWPE components for total joint arthroplasty?Biomed Res Int. 2014;2014:581812. doi: 10.1155/2014/581812. Epub 2014 Sep 11. Biomed Res Int. 2014. PMID: 25295264 Free PMC article.
-
Contact mechanics of modular metal-on-polyethylene total hip replacement under adverse edge loading conditions.J Biomech. 2014 Oct 17;47(13):3303-9. doi: 10.1016/j.jbiomech.2014.08.015. Epub 2014 Sep 1. J Biomech. 2014. PMID: 25218504 Free PMC article.
-
Wear resistance and mechanical properties of highly cross-linked, ultrahigh-molecular weight polyethylene doped with vitamin E.J Arthroplasty. 2006 Jun;21(4):580-91. doi: 10.1016/j.arth.2005.07.009. J Arthroplasty. 2006. PMID: 16781413 Free PMC article.
-
Improved resistance to wear and fatigue fracture in high pressure crystallized vitamin E-containing ultra-high molecular weight polyethylene.Biomaterials. 2009 Apr;30(10):1870-80. doi: 10.1016/j.biomaterials.2008.12.029. Epub 2009 Jan 8. Biomaterials. 2009. PMID: 19135247 Free PMC article.