Initial clinical experience with a video-based patient positioning system
- PMID: 10477025
- DOI: 10.1016/s0360-3016(99)00182-0
Initial clinical experience with a video-based patient positioning system
Abstract
Purpose: To report initial clinical experience with an interactive, video-based patient positioning system that is inexpensive, quick, accurate, and easy to use.
Methods and materials: System hardware includes two black-and-white CCD cameras, zoom lenses, and a PC equipped with a frame grabber. Custom software is used to acquire and archive video images, as well as to display real-time subtraction images revealing patient misalignment in multiple views. Two studies are described. In the first study, video is used to document the daily setup histories of 5 head and neck patients. Time-lapse cine loops are generated for each patient and used to diagnose and correct common setup errors. In the second study, 6 twice-daily (BID) head and neck patients are positioned according to the following protocol: at AM setups conventional treatment room lasers are used; at PM setups lasers are used initially and then video is used for 1-2 minutes to fine-tune the patient position. Lateral video images and lateral verification films are registered off-line to compare the distribution of setup errors per patient, with and without video assistance.
Results: In the first study, video images were used to determine the accuracy of our conventional head and neck setup technique, i.e., alignment of lightcast marks and surface anatomy to treatment room lasers and the light field. For this initial cohort of patients, errors ranged from sigma = 5 to 7 mm and were patient-specific. Time-lapse cine loops of the images revealed sources of the error, and as a result, our localization techniques and immobilization device were modified to improve setup accuracy. After the improvements, conventional setup errors were reduced to sigma = 3 to 5 mm. In the second study, when a stereo pair of live subtraction images were introduced to perform daily "on-line" setup correction, errors were reduced to sigma = 1 to 3 mm. Results depended on patient health and cooperation and the length of time spent fine-tuning the position.
Conclusion: An interactive, video-based patient positioning system was shown to reduce setup errors to within 1 to 3 mm in head and neck patients, without a significant increase in overall treatment time or labor-intensive procedures. Unlike retrospective portal image analysis, use of two live-video images provides the therapists with immediate feedback and allows for true 3-D positioning and correction of out-of-plane rotation before radiation is delivered. With significant improvement in head and neck alignment and the elimination of setup errors greater than 3 to 5 mm, margins associated with treatment volumes potentially can be reduced, thereby decreasing normal tissue irradiation.
Similar articles
-
Performance of a video-image-subtraction-based patient positioning system.Int J Radiat Oncol Biol Phys. 1997 Jul 1;38(4):855-66. doi: 10.1016/s0360-3016(97)00081-3. Int J Radiat Oncol Biol Phys. 1997. PMID: 9240655
-
The impact of daily setup variations on head-and-neck intensity-modulated radiation therapy.Int J Radiat Oncol Biol Phys. 2005 Mar 1;61(3):779-88. doi: 10.1016/j.ijrobp.2004.07.696. Int J Radiat Oncol Biol Phys. 2005. PMID: 15708257 Clinical Trial.
-
Portal imaging protocol for radical dose-escalated radiotherapy treatment of prostate cancer.Int J Radiat Oncol Biol Phys. 1998 Jan 1;40(1):221-31. doi: 10.1016/s0360-3016(97)00551-8. Int J Radiat Oncol Biol Phys. 1998. PMID: 9422580 Clinical Trial.
-
Comparison of setup accuracy of three different thermoplastic masks for the treatment of brain and head and neck tumors.Radiother Oncol. 2001 Feb;58(2):155-62. doi: 10.1016/s0167-8140(00)00280-2. Radiother Oncol. 2001. PMID: 11166866
-
Effect of patient setup errors on simultaneously integrated boost head and neck IMRT treatment plans.Int J Radiat Oncol Biol Phys. 2005 Oct 1;63(2):422-33. doi: 10.1016/j.ijrobp.2005.02.029. Int J Radiat Oncol Biol Phys. 2005. PMID: 16168835
Cited by
-
Comparison of surface matching and target matching for image-guided pelvic radiation therapy for both supine and prone patient positions.J Appl Clin Med Phys. 2016 May 8;17(3):14-24. doi: 10.1120/jacmp.v17i3.5611. J Appl Clin Med Phys. 2016. PMID: 27167254 Free PMC article.
-
Verification of quality parameters for portal images in radiotherapy.Radiol Oncol. 2011 Mar;45(1):68-74. doi: 10.2478/v10019-010-0052-6. Epub 2010 Dec 31. Radiol Oncol. 2011. PMID: 22933938 Free PMC article.
-
A survey of image-guided radiation therapy use in the United States.Cancer. 2010 Aug 15;116(16):3953-60. doi: 10.1002/cncr.25129. Cancer. 2010. PMID: 20564090 Free PMC article.
-
Target registration errors with surface imaging system in conformal radiotherapy for prostate cancer: study on 19 patients.Radiol Med. 2012 Dec;117(8):1419-28. doi: 10.1007/s11547-012-0823-9. Epub 2012 May 14. Radiol Med. 2012. PMID: 22580804 English, Italian.
-
Clinical results of a pilot study on stereovision-guided stereotactic radiotherapy and intensity modulated radiotherapy.Technol Cancer Res Treat. 2010 Dec;9(6):603-17. doi: 10.1177/153303461000900609. Technol Cancer Res Treat. 2010. PMID: 21070083 Free PMC article. Clinical Trial.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical