A new Lp(a) assay that is unaffected by apo(a) size polymorphism
- PMID: 10509894
- DOI: 10.1016/s0009-8981(99)00119-9
A new Lp(a) assay that is unaffected by apo(a) size polymorphism
Erratum in
- Clin Chim Acta 2000 Feb 25;292(1-2):187
Abstract
We developed sandwich ELISA methods in which anti-apo(a) kringle 4 type 5 through protease (K4 x 5-Pro) domain monoclonal antibody (clone: 203E2) is employed in each instance as the capture antibody and one of the three species of monoclonal antibody [Mab] (clones: 108B8, 202A9, 2B3) is used as the labeled antibody. Using serum containing apo(a) with 34 repeats of kringle 4 as the calibrator, a commercial kit using anti-Lp(a) polyclonal antibody (Pab) or anti-apo(a) Mab overestimated the Lp(a) concentration in samples containing apo(a) with more than 34 repeats of kringle 4 and underestimated the Lp(a) concentration in samples containing apo(a) with fewer than 34 repeats of kringle 4. Moreover, it was demonstrated that the ratios of commercial kit values to anti-apo(a) K4 x 5-Pro Mab-based method values increased as the size of apo(a) increased. The ratios of apo(a) K5 x Pro Mab-based method values to anti-apo(a) K4 x 5-Pro Mab-based method values, however, remained almost constant regardless of the size polymorphism. Thus, we suggest that apo(a) size heterogeneity can significantly affect Lp(a) measurement in the Lp(a) assay using anti-Lp(a) Pab. The novel Lp(a) assay method, using only anti-apo(a) K4 x 5-Pro Mab, is not subject to this phenomenon.