Motion at the modular acetabular shell and liner interface. A comparative study
- PMID: 10546629
Motion at the modular acetabular shell and liner interface. A comparative study
Abstract
Excessive polyethylene debris generated from a total hip arthroplasty can lead to osteolysis and premature revision. Most of this polyethylene debris comes from the concave articulation. However, abrasive wear on the convex side of a modular polyethylene component also may play a role in this problem. Motion of a modular polyethylene liner with respect to its shell can lead to such abrasive wear. Six samples of modular acetabular components from eight manufacturers were tested for motion between the shell and the liner. Motion at the shell liner interface was detected by five fiberoptic sensors in the x, y, and z planes. Micromotion occurred at the shell liner interface in all designs tested. The magnitude of motion varied between manufacturers, ranging from 5 to 311 microns. Design features that limit motion may be advantageous in limiting the amount of polyethylene debris.
Similar articles
-
Polyethylene wear debris in modular acetabular prostheses.Clin Orthop Relat Res. 1995 Aug;(317):44-56. Clin Orthop Relat Res. 1995. PMID: 7671495
-
The acetabular insert-metal backing interface: an additional source of polyethylene wear debris.J Arthroplasty. 2005 Oct;20(7):914-22. doi: 10.1016/j.arth.2005.04.033. J Arthroplasty. 2005. PMID: 16230245
-
Study of micromotion in modular acetabular components during gait and subluxation: a finite element investigation.J Biomech Eng. 2008 Apr;130(2):021002. doi: 10.1115/1.2898715. J Biomech Eng. 2008. PMID: 18412489
-
[Research progress of backside wear in acetabular liners].Zhongguo Xiu Fu Chong Jian Wai Ke Za Zhi. 2013 Dec;27(12):1453-6. Zhongguo Xiu Fu Chong Jian Wai Ke Za Zhi. 2013. PMID: 24640364 Review. Chinese.
-
Review of the evolution of the cementless acetabular cup.Orthopedics. 2008 Dec;31(12 Suppl 2):orthosupersite.com/view.asp?rID=37178. Orthopedics. 2008. PMID: 19298028 Review.
Cited by
-
Early follow-up for a hybrid total hip arthroplasty using a metal-backed acetabular component designed to reduce "backside" polyethylene wear.HSS J. 2005 Sep;1(1):31-4. doi: 10.1007/s11420-005-0102-6. HSS J. 2005. PMID: 18751806 Free PMC article.
-
Direct comparison of polyethylene wear in cemented and uncemented acetabular cups.J Orthop Traumatol. 2010 Sep;11(3):155-8. doi: 10.1007/s10195-010-0104-0. Epub 2010 Sep 2. J Orthop Traumatol. 2010. PMID: 20811923 Free PMC article.
-
Survivorship of monoblock trabecular metal cups in primary THA : midterm results.Clin Orthop Relat Res. 2008 Jan;466(1):159-66. doi: 10.1007/s11999-007-0008-3. Epub 2008 Jan 3. Clin Orthop Relat Res. 2008. PMID: 18196389 Free PMC article.
-
Aseptic loosening of total hip replacement presenting as an anterior thigh mass.Musculoskelet Surg. 2013 Dec;97(3):247-9. doi: 10.1007/s12306-011-0167-y. Epub 2011 Oct 9. Musculoskelet Surg. 2013. PMID: 21984313
-
Backside Wear Is Not Dependent on the Acetabular Socket Design in Crosslinked Polyethylene Liners.Clin Orthop Relat Res. 2016 Feb;474(2):374-82. doi: 10.1007/s11999-015-4471-y. Clin Orthop Relat Res. 2016. PMID: 26201422 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources