Comparative Study
. 1999 Oct;9(1-2):42-51.
doi: 10.1007/BF02634591.

T2-weighted breathold imaging of the liver: a quantitative and qualitative comparison of fast spin echo and half Fourier single shot fast spin echo imaging

Affiliations
Comparative Study

T2-weighted breathold imaging of the liver: a quantitative and qualitative comparison of fast spin echo and half Fourier single shot fast spin echo imaging

T K Helmberger et al. MAGMA. 1999 Oct.

Abstract

The imaging characteristics of two EPI-hybrid breath-hold sequences, T2-weighted fast spin-echo [FSE, effective echo time (TEeff) 138 ms] and half Fourier single shot turbo spin-echo (HASTE, TEeff 60 ms), were compared in hepatic imaging. A total of 111 patients with suspected hepatic disease were studied at 1.5 Tesla using a body phase-array coil. The signal-to-noise (S/N) and contrast-to-noise (C/N) ratios for organs and lesions were calculated and quantitatively compared. Organ delineation, visualization of anatomical structures and pathological lesions, artifacts, and total image quality were qualitatively assessed and statistically compared. The final diagnoses were metastases from colorectal, breast, and pancreatic cancer in 23/111, hepatocellular carcinoma in 15/111, cysts in 19/111, hemangiomas in 9/111, several other lesions in 7/111, and no lesions in 38/111 of the cases. A total of 139 lesion in 73% of the patients were seen while 85% of the lesions were at least 1.5 cm in size. Regarding S/Ns HASTE was significantly (P < 0.03) superior to FSE with only minor (P > 0.05) differences in C/Ns between the two sequences for anatomical and pathological structures. HASTE demonstrated in almost all (97.3%) of the cases no artifacts, while on fast SE imaging moderate to minor artifacts were present in 23.5-51.7% of the cases. The overall image quality and diagnostic confidence was rated significantly higher (good 43.2%, excellent 53.2%) for HASTE than for fast SE imaging (good 44.8%, excellent 17.6%). Providing comparable C/Ns for anatomical and pathological structures, breatheld HASTE imaging proved to be superior to fast SE in T2-weighted imaging of the upper abdomen regarding general image quality, and, with adequate technical prerequisites, may be a suitable substitute of fast T2-imaging techniques.

PubMed Disclaimer

Similar articles

Cited by

References

    1. J Magn Reson Imaging. 1998 Mar-Apr;8(2):384-90 - PubMed
    1. Radiology. 1993 Dec;189(3):863-70 - PubMed
    1. AJR Am J Roentgenol. 1994 Dec;163(6):1389-93 - PubMed
    1. J Magn Reson Imaging. 1994 Jul-Aug;4(4):577-84 - PubMed
    1. Radiology. 1995 Jun;195(3):769-76 - PubMed

Publication types

LinkOut - more resources