The AAPM/RSNA physics tutorial for residents. MR imaging safety considerations. Radiological Society of North America
- PMID: 10555679
- DOI: 10.1148/radiographics.19.6.g99no331641
The AAPM/RSNA physics tutorial for residents. MR imaging safety considerations. Radiological Society of North America
Abstract
Experience and research over the past decade have demonstrated that diagnostic magnetic resonance (MR) imaging is a biologically safe imaging modality. Specifically, there is currently no convincing evidence that there is any long-term or irreversible biologic effects associated with the radiation and magnetic fields used in MR imaging, specifically radio-frequency (RF) radiation, static magnetic fields, and time-varying gradient fields. However, numerous hazards of MR imaging do exist that can cause severe injuries or even death. These hazards are primarily the result of (a) strong magnetic fields and the strong force that they exert on ferromagnetic objects brought into their influence, including interference with electronic devices such as pacemakers and other implanted electronic devices, and (b) RF burns resulting from inadvertently induced currents in conductive loops placed on the patient's skin surface (eg, electrocardiographic leads and other monitoring devices). Other potential concerns are peripheral nerve stimulation resulting from rapidly switched gradients and auditory noise levels. Establishing a complete and coordinated educational program for all MR imaging facility personnel and conducting effective screening and preparation of patients scheduled for MR imaging procedures are essential to avoid accidents and RF burns and to maintain a safe MR imaging facility.
Similar articles
-
Safety considerations in MR imaging.Radiology. 1990 Sep;176(3):593-606. doi: 10.1148/radiology.176.3.2202008. Radiology. 1990. PMID: 2202008 Review.
-
Magnetic resonance safety.Methods Mol Biol. 2011;711:17-28. doi: 10.1007/978-1-61737-992-5_2. Methods Mol Biol. 2011. PMID: 21279596
-
AAPM/RSNA physics tutorial for residents: MR artifacts, safety, and quality control.Radiographics. 2006 Jan-Feb;26(1):275-97. doi: 10.1148/rg.261055134. Radiographics. 2006. PMID: 16418258 Review.
-
MR procedures: biologic effects, safety, and patient care.Radiology. 2004 Sep;232(3):635-52. doi: 10.1148/radiol.2323030830. Epub 2004 Jul 29. Radiology. 2004. PMID: 15284433 Review.
-
Safety aspects of radiofrequency power deposition in magnetic resonance.Magn Reson Imaging Clin N Am. 1998 Nov;6(4):775-89. Magn Reson Imaging Clin N Am. 1998. PMID: 9799855 Review.
Cited by
-
Muskuloskeletal MR imaging at 3.0 T: current status and future perspectives.Eur Radiol. 2006 Jun;16(6):1298-307. doi: 10.1007/s00330-006-0184-7. Epub 2006 Mar 16. Eur Radiol. 2006. PMID: 16541224 Review.
-
Effects of external electrical and magnetic fields on pacemakers and defibrillators: from engineering principles to clinical practice.Circulation. 2013 Dec 24;128(25):2799-809. doi: 10.1161/CIRCULATIONAHA.113.005697. Circulation. 2013. PMID: 24366589 Free PMC article.
-
Patient Monitoring System for MRI: An Evidence-Based Analysis.Ont Health Technol Assess Ser. 2003;3(7):1-16. Epub 2003 Dec 1. Ont Health Technol Assess Ser. 2003. PMID: 23074447 Free PMC article. No abstract available.
-
[MRT-induced burns in tattooed patients. Case report of an traumatic surgery patient].Unfallchirurg. 2005 May;108(5):410-4. doi: 10.1007/s00113-004-0877-9. Unfallchirurg. 2005. PMID: 15909207 German.
-
Controlled E-field gradient coils.MAGMA. 2003 Nov;16(3):113-20. doi: 10.1007/s10334-003-0015-7. Epub 2003 Oct 31. MAGMA. 2003. PMID: 14593514
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical
Miscellaneous