Assessment of the reproducibility of baseline and hyperemic myocardial blood flow measurements with 15O-labeled water and PET
- PMID: 10565780
Assessment of the reproducibility of baseline and hyperemic myocardial blood flow measurements with 15O-labeled water and PET
Abstract
PET with 15O-labeled water allows noninvasive quantification of myocardial blood flow (MBF) at baseline and during pharmacologically induced hyperemia to assess the coronary vasodilator reserve (CVR = hyperemic/baseline MBF). Despite widespread use of PET, its reproducibility during one study session has not been tested. Intravenous adenosine (Ado), a powerful coronary vasodilator with a very short decay time, is commonly used for the induction of hyperemia. However, it is not known whether Ado can induce tachyphylaxis after short-term repetitive administration. In this study, we aimed to test the reproducibility of PET assessment of CVR during Ado-induced hyperemia.
Methods: In 21 healthy volunteer men, baseline and Ado MBF were measured twice using PET with 15O-labeled water to obtain two CVR assessments within 1 h.
Results: There was no significant difference between the two baselines (0.89 +/- 0.14 versus 0.99 +/- 0.15 mL/min/g, mean difference 13% +/- 11%) or between the two hyperemic MBFs (3.51 +/- 0.45 versus 3.83 +/- 0.49 mL/min/g, mean difference 10% +/- 14%), resulting in comparable values of CVR (4.05 +/- 0.75 versus 3.93 +/- 0.72, mean difference 2% +/- 15%). The repeatability coefficient for MBF was 0.17 mL/min/g at baseline and 0.94 mL/min/g during hyperemia. The repeatability coefficient of the rate pressure product (RPP) was lower at baseline (1,304 mm Hg x beat/min) than during hyperemia (3,448 mm Hg x beat/min).
Conclusion: Repeated measurements of MBF and CVR during the same study session were not significantly different, demonstrating the validity of the technique. The larger variability of hyperemic flow, as indicated by the larger repeatability coefficient, was paralleled by a greater variability of the RPP. This could mean that the greater variability of MBF during stress is more likely due to a variable response to Ado rather than to a measurement error.
Similar articles
-
Quantitative relation between hemodynamic changes during intravenous adenosine infusion and the magnitude of coronary hyperemia: implications for myocardial perfusion imaging.J Am Coll Cardiol. 2005 Feb 15;45(4):553-8. doi: 10.1016/j.jacc.2004.10.064. J Am Coll Cardiol. 2005. PMID: 15708703
-
Effect of caffeine intake on myocardial hyperemic flow induced by adenosine triphosphate and dipyridamole.J Nucl Med. 2004 May;45(5):730-8. J Nucl Med. 2004. PMID: 15136619
-
Reproducibility of measurements of regional myocardial blood flow in a model of coronary artery disease: Comparison of H215O and 13NH3 PET techniques.J Nucl Med. 2006 Jul;47(7):1193-201. J Nucl Med. 2006. PMID: 16818955
-
Bicycle exercise stress in PET for assessment of coronary flow reserve: repeatability and comparison with adenosine stress.J Nucl Med. 2003 Feb;44(2):146-54. J Nucl Med. 2003. PMID: 12571202 Clinical Trial.
-
Repeatability of rest and hyperemic myocardial blood flow measurements with 82Rb dynamic PET.J Nucl Med. 2009 Jan;50(1):68-71. doi: 10.2967/jnumed.108.055673. Epub 2008 Dec 17. J Nucl Med. 2009. PMID: 19091892 Clinical Trial.
Cited by
-
Comparison of myocardial blood flow and flow reserve with dobutamine and dipyridamole stress using rubidium-82 positron emission tomography.J Nucl Cardiol. 2021 Feb;28(1):34-45. doi: 10.1007/s12350-020-02186-1. Epub 2020 May 24. J Nucl Cardiol. 2021. PMID: 32449001
-
Evaluation of quantitative CMR perfusion imaging by comparison with simultaneous 15O-water-PET.J Nucl Cardiol. 2021 Aug;28(4):1252-1266. doi: 10.1007/s12350-019-01810-z. Epub 2019 Jul 16. J Nucl Cardiol. 2021. PMID: 31313066 Free PMC article.
-
Feasibility and operator variability of myocardial blood flow and reserve measurements with ⁹⁹mTc-sestamibi quantitative dynamic SPECT/CT imaging.J Nucl Cardiol. 2014 Dec;21(6):1075-88. doi: 10.1007/s12350-014-9971-8. Epub 2014 Oct 4. J Nucl Cardiol. 2014. PMID: 25280761 Clinical Trial.
-
Quantitative myocardial blood flow imaging with integrated time-of-flight PET-MR.EJNMMI Phys. 2017 Dec;4(1):1. doi: 10.1186/s40658-016-0171-2. Epub 2017 Jan 6. EJNMMI Phys. 2017. PMID: 28058674 Free PMC article.
-
Segmental quantitative myocardial perfusion with PET for the detection of significant coronary artery disease in patients with stable angina.Eur J Nucl Med Mol Imaging. 2016 Jul;43(8):1522-9. doi: 10.1007/s00259-016-3362-0. Epub 2016 Mar 18. Eur J Nucl Med Mol Imaging. 2016. PMID: 26993310
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources