Noninvasive glucose monitoring: comprehensive clinical results. Cygnus Research Team
- PMID: 10573275
- DOI: 10.1001/jama.282.19.1839
Noninvasive glucose monitoring: comprehensive clinical results. Cygnus Research Team
Abstract
Context: Intensive diabetes management using frequent blood glucose measurements to guide therapy has been shown to significantly improve short- and long-term outcomes. Development of a device that makes possible frequent, automatic, painless, and accurate measurements of glucose would facilitate intensive management.
Objective: To determine the accuracy of the GlucoWatch automatic glucose biographer (Cygnus Inc) compared with that of serial blood glucose measurements.
Design: Multicenter comparative study of the GlucoWatch biographer and the HemoCue blood glucose analyzer (Aktiebolaget Leo) performed between August 29 and October 17, 1998. Participants wore up to 2 biographers during the 15-hour study session and performed 2 fingersticks per hour for comparative blood glucose measurements. The biographers were calibrated with a single HemoCue measurement after a 3-hour warm-up period. Diet and insulin were manipulated to produce a broad glycemic range during the study.
Setting: Controlled clinical environment at 2 diabetes centers and 3 contract research organizations in the United States.
Participants: A total of 92 subjects (mean [SD] age, 42.1 [15.1] years; 59.8% women) with type 1 or 2 diabetes requiring treatment with insulin.
Main outcome measures: Mean error, mean absolute error, correlation, slope, and intercept using Deming regression, and clinical significance of differences between biographer readings and blood glucose measurements using the Clarke error grid.
Results: Results showed close tracking of blood glucose over a range of 2.2 to 22.2 mmol/L (40-400 mg/dL) for up to 12 hours using a single point calibration. The biographer readings lagged behind serial blood glucose values by a mean of 18 minutes. An analysis of 2167 data pairs shows a linear relationship (r = 0.88; slope = 1.03; intercept = -0.33 mmol/L [-6 mg/dL]) between biographer readings and serial glucose measurements. The mean absolute error between the 2 measurements was 15.6% (mean error [SD], -0.07 [1.82] mmol/L [-1 [33] mg/dL]), and 96.8% of the data fell in the therapeutically relevant regions of the error grid analysis.
Conclusion: These results demonstrate close agreement between GlucoWatch biographer readings and blood glucose measurements using repeated fingerstick blood samples. The automatic, frequent, and noninvasive measurements obtained with the biographer provides more information about glucose levels than the current standard of care.
Similar articles
-
Correlation of fingerstick blood glucose measurements with GlucoWatch biographer glucose results in young subjects with type 1 diabetes.Diabetes Care. 1999 Oct;22(10):1708-14. doi: 10.2337/diacare.22.10.1708. Diabetes Care. 1999. PMID: 10526740 Clinical Trial.
-
Glucose monitoring by reverse iontophoresis.Diabetes Metab Res Rev. 2002 Jan-Feb;18 Suppl 1:S49-53. doi: 10.1002/dmrr.210. Diabetes Metab Res Rev. 2002. PMID: 11921430 Review.
-
Effect of acetaminophen on the accuracy of glucose measurements obtained with the GlucoWatch biographer.Diabetes Technol Ther. 2000 Summer;2(2):199-207. doi: 10.1089/15209150050025140. Diabetes Technol Ther. 2000. PMID: 11469259
-
The GlucoWatch biographer: a frequent automatic and noninvasive glucose monitor.Ann Med. 2000 Dec;32(9):632-41. doi: 10.3109/07853890009002034. Ann Med. 2000. PMID: 11209971 Review.
-
Use of the Cygnus GlucoWatch biographer at a diabetes camp.Pediatrics. 2004 Jan;113(1 Pt 1):108-11. doi: 10.1542/peds.113.1.108. Pediatrics. 2004. PMID: 14702457 Clinical Trial.
Cited by
-
Evolving approaches to intensive insulin therapy in type 1 diabetes: multiple daily injections, insulin pumps and new methods of monitoring.Rev Endocr Metab Disord. 2003 Dec;4(4):325-34. doi: 10.1023/a:1027345927676. Rev Endocr Metab Disord. 2003. PMID: 14618017 Review. No abstract available.
-
Advances in diabetes for the millennium: insulin treatment and glucose monitoring.MedGenMed. 2004 Sep 16;6(3 Suppl):8. MedGenMed. 2004. PMID: 15647713 Free PMC article. Review.
-
Simultaneous transdermal extraction of glucose and lactate from human subjects by reverse iontophoresis.Int J Nanomedicine. 2008;3(2):211-23. doi: 10.2147/ijn.s1728. Int J Nanomedicine. 2008. PMID: 18686780 Free PMC article.
-
A novel multipurpose device for dataset creation and on-device immediate estimation of blood glucose level from reflection ppg.Heliyon. 2023 Sep 1;9(9):e19553. doi: 10.1016/j.heliyon.2023.e19553. eCollection 2023 Sep. Heliyon. 2023. PMID: 37810055 Free PMC article.
-
Flash Glucose Monitoring: The Future Is Here.Diabetes Technol Ther. 2017 May;19(S2):S1-S3. doi: 10.1089/dia.2017.0098. Diabetes Technol Ther. 2017. PMID: 28541133 Free PMC article. No abstract available.