Continuous cardiac output by femoral arterial thermodilution calibrated pulse contour analysis: comparison with pulmonary arterial thermodilution

Affiliations
Comparative Study

Continuous cardiac output by femoral arterial thermodilution calibrated pulse contour analysis: comparison with pulmonary arterial thermodilution

O Goedje et al. Crit Care Med. 1999 Nov.

Abstract

Objective: To compare two thermodilution methods for the determination of cardiac output (CO)-thermodilution in the pulmonary artery (COpa) and thermodilution in the femoral artery (COa)-with each other and with CO determined by continuous pulse contour analysis (COpc) in terms of reproducibility, bias, and correlation among the different methods. Good agreement between the methods would indicate the potential of pulse contour analysis to monitor CO continuously and at reduced invasiveness.

Design: Prospective criterion standard study.

Setting: Cardiac surgical intensive care unit in a university hospital.

Patients: Twenty-four postoperative cardiac surgery patients.

Interventions: Without interfering with standard hospital cardiac recovery procedures, changes in CO as a result of the postsurgical course, administration of vasoactive substances, and/or fluid administration were recorded. CO was first recorded after a 1-hr stabilization period in the intensive care unit and hourly thereafter for 6 hrs, and by subsequent determinations at 9, 12, and 24 hrs.

Measurements and main results: There were 216 simultaneous determinations of COpa, COa, and COpc. COpc was initially calibrated using COa, and no further recalibration of COpc was performed. COpa ranged from 3.0 to 11.8 L/min, and systemic vascular resistance ranged from 252 to 2434 dyne x sec/cm5. The mean difference (bias) +/-2 SD of differences (limits of agreement) was -0.29+/-1.31 L/min for COpa vs. COa, 0.07+/-1.4 L/min for COpc vs. COpa, and -0.22+/-1.58 L/min for COpc vs. COa. In all but four patients COpc correlated with COa after the initial calibration. Correlation and precision of COpc vs. COa was stable for 24 hrs.

Conclusions: Femoral artery pulse contour CO correlates well with both COpa and COa even during substantial variations in vascular tone and hemodynamics. Additionally, CO determined by arterial thermodilution correlates well with COpa. Thus, COa can be used to calibrate COpc.