Accuracy of T2-weighted fast spin-echo MR imaging with fat saturation in detecting cartilage defects in the knee: comparison with arthroscopy in 130 patients

Affiliations
Comparative Study

Accuracy of T2-weighted fast spin-echo MR imaging with fat saturation in detecting cartilage defects in the knee: comparison with arthroscopy in 130 patients

M A Bredella et al. AJR Am J Roentgenol. 1999 Apr.

Abstract

Objective: The purpose of this study was to assess the accuracy of routine T2-weighted MR imaging in detecting and grading articular cartilage lesions in the knee compared with arthroscopy.

Subjects and methods: We examined 130 consecutive patients who underwent MR imaging and arthroscopy of the knee for suspected internal derangement. MR imaging consisted of axial and coronal T2-weighted fast spin-echo sequences with fat saturation and sagittal T2-weighted spin-echo sequences. Each single plane was evaluated and graded for the presence and appearance of articular cartilage defects using a standard arthroscopic grading scheme adapted to MR imaging.

Results: Of the 86 arthroscopically proven abnormalities, 81 were detected on MR imaging. Sensitivity of the T2-weighted fast spin-echo sequence with fat saturation was 61% for the coronal plane alone and 59% for the axial plane alone. Specificity for each plane was 99%. Sensitivity for the sagittal T2-weighted spin-echo sequence was 40%, and specificity was 100%. Sensitivity of the combination of axial and coronal T2-weighted fast spin-echo sequences with fat saturation and sagittal T2-weighted spin-echo sequence compared with arthroscopy for revealing cartilage lesions was 94%, specificity was 99%, and accuracy was 98%. Sensitivity of coronal and axial T2-weighted fast spin-echo sequences with fat saturation was 93%, and specificity was 99%. Fifty-five lesions (64%) were identically graded on MR imaging and arthroscopy. Seventy-eight lesions (90%) were within one grade using MR imaging and arthroscopy, and 84 lesions (97%) were within two grades using MR imaging and arthroscopy.

Conclusion: T2-weighted fast spin-echo MR imaging with fat saturation is an accurate and fast technique for detecting and grading articular cartilage defects in the knee. The combination of the axial and coronal planes offers sufficient coverage of articular surfaces to provide a high sensitivity and specificity for chondral defects.