Dose reduction in CT by anatomically adapted tube current modulation. I. Simulation studies

Affiliations

Dose reduction in CT by anatomically adapted tube current modulation. I. Simulation studies

M Gies et al. Med Phys. 1999 Nov.

Abstract

Tube current modulation governed by x-ray attenuation during CT (computed tomography) acquisition can lead to noise reduction which in turn can be used to achieve patient dose reduction without loss in image quality. The potential of this technique was investigated in simulation studies calculating both noise amplitude levels and noise distribution in CT images. The dependence of noise on the inodulation function, amplitude of modulation, shape and size of the object, and possible phase shift between attenuation and modulation function were examined. Both sinusoidal and attenuation-based control functions were used to modulate tube current. Noise reduction was calculated for both ideal systems and for real systems with limited modulation amplitude. Dose reductions up to 50% can be achieved depending on the phantom geometry and tube current modulation function. Attenuation-based tube current modulation yields substantially higher reduction than fixed-shape modulation functions. Optimal results are obtained when the current is modulated as a function of the square root of attenuation. A modulation amplitude of at least 90% should be available to exploit the potential of these techniques.

PubMed Disclaimer

Substances