Gelfoam as a barrier to prevent polymethylmethacrylate-induced thermal injury of the spinal cord: in vitro and in vivo studies in pigs
- PMID: 10598992
Gelfoam as a barrier to prevent polymethylmethacrylate-induced thermal injury of the spinal cord: in vitro and in vivo studies in pigs
Abstract
Gelatin sponge (Gelfoam; Upjohn, Kalamazoo, MI, U.S.A.) is commonly used as an interpositional barrier to shield the spinal cord from thermal injury during vertebral reconstruction with polymethylmethacrylate bone cement. The aim of this study was to record epidural and intradural temperatures during polymethylmethacrylate reconstruction of vertebral corpectomy defects. Three surgical techniques (subtotal corpectomy, total corpectomy with insertion of a Gelfoam barrier, and total corpectomy with no barrier) were compared in vivo and in vitro in a porcine model. As expected, total corpectomy defects cemented without a Gelfoam barrier produced the highest epidural temperatures in vivo (52.8 degrees C) and in vitro (58 +/- 2 degrees C). The Gelfoam barrier provided some protection against heat transfer, but peak temperatures and absolute temperature increases were significantly higher than in defects with an intact posterior cortex (p < 0.05). These results indicate that an intact posterior cortex provides the best protection against heat transfer, whereas the use of a Gelfoam barrier appears to provide only partial protection against thermal injury.
Similar articles
-
The safety of cement fixation in the cervical spine. Studies of a rabbit model.Clin Orthop Relat Res. 1979 Mar-Apr;(139):276-82. Clin Orthop Relat Res. 1979. PMID: 455844
-
Temperature measurement during polymerization of polymethylmethacrylate cement used for vertebroplasty.Spine (Phila Pa 1976). 2003 Jul 15;28(14):1555-9. Spine (Phila Pa 1976). 2003. PMID: 12865844
-
In vivo temperature profile of intervertebral discs and vertebral endplates during vertebroplasty: an experimental study in sheep.Spine (Phila Pa 1976). 2006 Jul 1;31(15):1674-8; discussion 1679. doi: 10.1097/01.brs.0000224193.52587.d8. Spine (Phila Pa 1976). 2006. PMID: 16816761
-
Combined extraforaminal and intradiscal cement leakage following percutaneous vertebroplasty.Spine (Phila Pa 1976). 2007 May 20;32(12):E358-62. doi: 10.1097/BRS.0b013e31805c0b25. Spine (Phila Pa 1976). 2007. PMID: 17515810 Review.
-
Renal cement embolism during percutaneous vertebroplasty.Eur Spine J. 2006 Oct;15 Suppl 5(Suppl 5):590-4. doi: 10.1007/s00586-005-0037-0. Epub 2005 Dec 14. Eur Spine J. 2006. PMID: 16362386 Free PMC article. Review.
Cited by
-
Modification of Mechanical Properties, Polymerization Temperature, and Handling Time of Polymethylmethacrylate Cement for Enhancing Applicability in Vertebroplasty.Biomed Res Int. 2016;2016:7901562. doi: 10.1155/2016/7901562. Epub 2016 Oct 12. Biomed Res Int. 2016. PMID: 27812530 Free PMC article.
-
In vitro investigation of heat transfer in calf spinal cord during polymethylmethacrylate application for vertebral body reconstruction.Eur Spine J. 2006 Mar;15(3):341-6. doi: 10.1007/s00586-004-0869-z. Epub 2005 May 24. Eur Spine J. 2006. PMID: 15912351 Free PMC article.
-
Effect of Percutaneous Kyphoplasty on the progression of intervertebral disc degeneration: a retrospective cohort study.J Orthop Surg Res. 2023 Mar 6;18(1):172. doi: 10.1186/s13018-023-03627-6. J Orthop Surg Res. 2023. PMID: 36879260 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical