Objectives: This pilot study was designed to assess the ability of a photoscreening camera to detect amblyogenic factors such as high refractive error, anisometropia, media opacities and strabismus in children, compared to the standard vision screening techniques employed by the local state public health screening personnel.
Methods: Public health personnel in Illinois used the Eyecor prototype to the current commercially available MTI PS-100 photoscreening camera (manufactured by Medical Technology Inc.) to screen 127 non-dilated subjects, ages 7 months to 20 years (mean age, 6 years), for amblyogenic factors. All participants were concurrently subjected to the "standard" vision screening employed routinely by state public health personnel. The study participants included a group of normal inner-city children and one group of special-needs children. The normal children came from both a public school and a private school. The special-needs children included a group of children from a state-run school for the deaf and hard-of-hearing and a separate group of children attending a multi-disciplinary Easter Seals clinic.
Results: In the population of normal children, the mean sensitivity and specificity for the observers using the Eyecor Camera system was 81% and 83% with a mean positive predictive value of 83%, and a mean negative predictive value of 55%. Standard vision screening techniques employed by public health service certified vision screeners had a sensitivity of 88%, a specificity of 91%, a positive predictive value of 67% and a negative predictive value of 97% in the same subjects. In the population of special-needs children with hearing impairment and developmental delay, the mean sensitivity and specificity for the observers using the Eyecor Camera system was 74% and 82% with a mean positive predictive value of 69% and a mean negative predictive value of 85%. Standard vision screening techniques employed by public health service certified vision screeners had a sensitivity of 100%, a specificity of 55%, a positive predictive value of 62% and a negative predictive value of 100% in the same subjects. (See Table 1)
Conclusions: This study shows that the Eyecor Photoscreening Camera is useful in screening normal children for amblyogenic factors. Photoscreening was at least as effective as standard screening methodologies performed by certified vision screeners as required by state public health policy. In addition, photoscreening is particularly useful in testing those children "unscreenable" by conventional vision screening procedures.
Similar articles
-
The efficacy of photoscreening for amblyopiagenic factors in a high risk population.Binocul Vis Strabismus Q. 2003;18(4):233-40. Binocul Vis Strabismus Q. 2003. PMID: 14653776
-
Screening for amblyopia in preverbal children with photoscreening photographs: IV. Interobserver variability in photograph grading: origin and method of reduction.Binocul Vis Strabismus Q. 2005;20(2):71-80. Binocul Vis Strabismus Q. 2005. PMID: 16042536
-
Screening for amblyogenic risk factors using the PlusoptiX S04 photoscreener on the indigent population of Honduras.Ophthalmology. 2010 Sep;117(9):1848-50. doi: 10.1016/j.ophtha.2010.01.038. Epub 2010 May 15. Ophthalmology. 2010. PMID: 20472292
-
Major review: The underutilization of vision screening (for amblyopia, optical anomalies and strabismus) among preschool age children.Binocul Vis Strabismus Q. 2003;18(4):217-32. Binocul Vis Strabismus Q. 2003. PMID: 14653775 Review.
-
Plusoptix photoscreener use for paediatric vision screening in Flanders and Iran.Acta Ophthalmol. 2020 Feb;98(1):80-88. doi: 10.1111/aos.14144. Epub 2019 Jun 14. Acta Ophthalmol. 2020. PMID: 31197950 Free PMC article. Review.
Cited by
-
Tests for detecting strabismus in children aged 1 to 6 years in the community.Cochrane Database Syst Rev. 2017 Nov 6;11(11):CD011221. doi: 10.1002/14651858.CD011221.pub2. Cochrane Database Syst Rev. 2017. PMID: 29105728 Free PMC article.
-
The value of serial personal photographs in timing the onset of unilateral cataracts in children.J AAPOS. 2009 Oct;13(5):459-62. doi: 10.1016/j.jaapos.2009.08.007. J AAPOS. 2009. PMID: 19840724 Free PMC article.
-
The relationship between anisometropia, patient age, and the development of amblyopia.Trans Am Ophthalmol Soc. 2005;103:313-36. Trans Am Ophthalmol Soc. 2005. PMID: 17057809 Free PMC article.
-
Canon CP-TX1 camera - As a screening tool for amblyogenic risk factors.Indian J Ophthalmol. 2022 Apr;70(4):1313-1316. doi: 10.4103/ijo.IJO_2161_21. Indian J Ophthalmol. 2022. PMID: 35326044 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Medical