. 2000 Jan;14(1):82-5.
doi: 10.1007/s004649900018.

Ultrasonic coagulator for video-assisted internal mammary artery harvest

Affiliations

Ultrasonic coagulator for video-assisted internal mammary artery harvest

T Ohtsuka et al. Surg Endosc. 2000 Jan.

Abstract

Background: The Harmonic Scalpel (HS; Ethicon Endo-Surgery, Cincinnati, OH, USA) is an ultrasonic coagulator that generates less heat than electrocautery. We compared canine internal mammary arteries (IMAs) harvested using the HS or electrocautery and reviewed the early clinical outcome after thoracoscopic IMA takedown with the HS.

Methods: Using HS and bipolar electrocautery (BE), 51 and 49 IMA branches, respectively, in six mongrel dogs were divided. The divided branches were subjected to a pressure tolerance test. The impact of coagulation on the main trunk was investigated microscopically. Thoracoscopic IMA harvesting with the HS was performed in 41 patients. Seven human IMA segments including 10 branches were examined microscopically. Postoperative angiography of 40 IMA grafts (97.6%) was performed.

Results: All the IMA branches withstood intraluminal pressures up to 250 mmHg. With BE, coagulation reached the adventitia of the main trunk in six segments (12.2%), but the HS caused no injury. Thoracoscopic IMA harvesting with excellent hemostasis was achieved using only the HS in all 41 patients. The main trunks of the human IMA segments were microscopically intact, and postoperative angiography demonstrated that the IMA grafts were intact.

Conclusion: The HS appears to be a reliable coagulator for thoracoscopic IMA harvesting with minimal damage.

PubMed Disclaimer

References

    1. J Am Coll Surg. 1995 Oct;181(4):347-51 - PubMed
    1. Ann Thorac Surg. 1998 Apr;65(4):984-5 - PubMed
    1. Ann Thorac Surg. 1987 Jun;43(6):676-7 - PubMed
    1. Endosc Surg Allied Technol. 1994 Jun-Aug;2(3-4):181-5 - PubMed
    1. Ann Thorac Surg. 1997 Jun;63(6 Suppl):S107-9 - PubMed

LinkOut - more resources