A retrospective study on perfusion incidents and safety devices
Abstract
Despite the acceptance of extracorporeal circulation as an effective modality to facilitate cardiac surgery, patient outcomes can be negatively influenced by the occurrence of perfusion incidents. A perfusion survey was conducted to identify safety techniques and incidents related to cardiopulmonary bypass (CPB). An 80-question survey was mailed to chief perfusionists of all 1030 USA cardiac surgical centers using CPB. The survey was designed to examine practices and incidents that occurred during a 2-year period (July 1996 to July 1998). Five-hundred-and-fifty-two (54% response rate) surveys were returned, which accounted for 797 hospitals (79% of all cardiac centers) and 653,621 surgical procedures. Of the 27 identified CPB safety devices, the highest utilization was arterial line filters (98.5%) and the lowest arterial line bubble traps (3.4%). Of the reported cases, a CPB incident occurred once every 138 cases. The most common occurring incidents were protamine reactions (1:783), coagulation problems (1:771), and heater/cooler failures 11:1809). The rate of occurrence of an incident resulting in a serious injury or death was one for every 1453 procedures. Although techniques and safety devices create a relatively secure environment for CPB, lower incident rates may be achieved with further improvements in coagulation monitoring and incident reporting.
Similar articles
-
Survey: retrospective survey of monitoring/safety devices and incidents of cardiopulmonary bypass for cardiac surgery in France.J Extra Corpor Technol. 2007 Sep;39(3):142-57; discussion 158-9. J Extra Corpor Technol. 2007. PMID: 17972449 Free PMC article.
-
An update on perfusion safety: does the type of perfusion practice affect the rate of incidents related to cardiopulmonary bypass?Perfusion. 2001 May;16(3):189-98. doi: 10.1177/026765910101600304. Perfusion. 2001. PMID: 11419654
-
Australasian perfusion incident survey.Perfusion. 1997 Sep;12(5):279-88. doi: 10.1177/026765919701200502. Perfusion. 1997. PMID: 9300472
-
Perfusion safety: defining the problem.Perfusion. 2005 Jul;20(4):195-203. doi: 10.1191/0267659105pf806oa. Perfusion. 2005. PMID: 16130365 Review.
-
[Patient safety in cardiopulmonary bypass; unusual problems and managements].Kyobu Geka. 2008 Jul;61(8 Suppl):636-9. Kyobu Geka. 2008. PMID: 20715401 Review. Japanese.
Cited by
-
Should Air Bubble Detectors Be Used to Quantify Microbubble Activity during Cardiopulmonary Bypass?J Extra Corpor Technol. 2015 Sep;47(3):174-9. J Extra Corpor Technol. 2015. PMID: 26543252 Free PMC article.
-
Survey: retrospective survey of monitoring/safety devices and incidents of cardiopulmonary bypass for cardiac surgery in France.J Extra Corpor Technol. 2007 Sep;39(3):142-57; discussion 158-9. J Extra Corpor Technol. 2007. PMID: 17972449 Free PMC article.
-
Preliminary analysis of perfusionists' strategies for managing routine and failure mode scenarios in cardiopulmonary bypass.J Extra Corpor Technol. 2007 Sep;39(3):160-7. J Extra Corpor Technol. 2007. PMID: 17972450 Free PMC article.
-
2024 EACTS/EACTAIC/EBCP Guidelines on cardiopulmonary bypass in adult cardiac surgery.Interdiscip Cardiovasc Thorac Surg. 2025 Feb 5;40(2):ivaf002. doi: 10.1093/icvts/ivaf002. Interdiscip Cardiovasc Thorac Surg. 2025. PMID: 39949317 Free PMC article. No abstract available.
-
Isoflurane cracks the polycarbonate connector of extra-corporeal circuit -A case report-.Korean J Anesthesiol. 2010 Mar;58(3):304-6. doi: 10.4097/kjae.2010.58.3.304. Epub 2010 Mar 29. Korean J Anesthesiol. 2010. PMID: 20498783 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources
Medical