Accuracy of oximetry with thermistor (OxiFlow) for diagnosis of obstructive sleep apnea and hypopnea

Affiliations
  • PMID: 10678466

Accuracy of oximetry with thermistor (OxiFlow) for diagnosis of obstructive sleep apnea and hypopnea

M A Baltzan et al. Sleep. .

Abstract

Objectives: To evaluate the diagnostic accuracy for obstructive sleep apnea and hypopnea (OSAH) of the OxiFlow (OF) device which combines oximetry with recording of thermistor airflow.

Design & setting: Patients scheduled for overnight diagnostic polysomnography (PSG) were studied with OF either simultaneously during laboratory PSG (L-OF, n=86), at home on a separate night (H-OF, n=66), or both (n=55).

Patients: 97 patients with suspected OSAH, of whom 40 had OSAH defined as an apnea-hypopnea index (AHI) of more than 15 events per hour of sleep on PSG.

Interventions: NA.

Measurements & results: The automated respiratory disturbance index (RDI) generated by the OF software considerably underestimated the AHI by PSG for both L-OF and H-OF. Altering the parameters for hypopnea identification by the software did not improve this. Visual inspection of the computerized OF tracings added considerable diagnostic information, but a manual count of RDI during visual review overestimated AHI. For the identification of cases vs. non-cases of OSAH, receiver operating characteristic area-under-the-curve statistics ranged from 0.77-0.90 for L-OF and from 0.71-0.77 for H-OF. Combining automated analysis with subsequent visual inspection of OF tracings yielded an overall sensitivity of 86% and specificity of 74% for the diagnosis of OSAH during H-OF recordings. Analysis of potential technician time saved indicated a benefit from the use of OF.

Conclusions: OF has diagnostic utility for the identification of OSAH. However, because of hardware and software limitations, it is unclear whether this device is superior to oximetry alone.