Effect of augmentation on the mechanics of vertebral wedge fractures
Abstract
Study design: The effect of cement augmentation of wedge-fractured vertebral bodies on spine segment compliance was studied in 16 cadaver specimens.
Objectives: 1) To assess the mechanical effects of cement augmentation of vertebral wedge fractures. 2) To determine whether a new reduction/injection procedure has the same mechanical effects as the established direct injection procedure.
Summary of background data: Although wedge fractures cause pain and disability in hundreds of thousands of people, few effective treatments are available. Clinical studies have shown that cement augmentation, a new procedure, effectively relieves pain and restores mobility in patients suffering from weak or fractured vertebrae. However, only a few studies have examined the mechanics of vertebral augmentation.
Methods: A wedge fracture was created in the middle vertebra of 16 three-vertebra cadaver spine segments. Neutral and full-load compliance of each fractured spine segment in flexion/extension and lateral bending were assessed by measuring the relative rotation of the vertebral bodies in response to applied moments. Eight of the fractured vertebral bodies were then augmented using direct injection, while the remaining eight fractured vertebral bodies were augmented using a combined reduction/injection procedure. Compliance of the augmented segments was then assessed.
Results: Augmentation significantly reduced the neutral compliance (reduction of 25% +/- 23%) (mean +/- standard deviation) and the full-load compliance (reduction of 23% +/- 20%) in flexion/extension (P < 0.005). Augmentation also significantly reduced the neutral compliance (reduction of 34% +/- 20%) and the full-load compliance (reduction of 26% +/- 17%) in lateral bending (P < 0.0001). No significant difference was found between the two procedures for compliance reduction.
Conclusions: Augmentation of wedge fractures using both direct injection and reduction/injection reduces spine segment compliance significantly.
Similar articles
-
Biomechanical changes after the augmentation of experimental osteoporotic vertebral compression fractures in the cadaveric thoracic spine.Spine J. 2005 Jan-Feb;5(1):55-63. doi: 10.1016/j.spinee.2004.08.005. Spine J. 2005. PMID: 15653085
-
Biomechanical comparison of kyphoplasty versus a titanium mesh implant with cement for stabilization of vertebral compression fractures.Spine (Phila Pa 1976). 2010 Sep 1;35(19):1783-8. doi: 10.1097/BRS.0b013e3181b7cc5d. Spine (Phila Pa 1976). 2010. PMID: 20098352
-
Long-term effects of vertebroplasty: adjacent vertebral fractures.J Long Term Eff Med Implants. 2006;16(4):265-80. doi: 10.1615/jlongtermeffmedimplants.v16.i4.10. J Long Term Eff Med Implants. 2006. PMID: 17073569 Review.
-
Vertebroplasty with high-viscosity polymethylmethacrylate cement facilitates vertebral body restoration in vitro.Spine (Phila Pa 1976). 2009 Nov 15;34(24):2619-25. doi: 10.1097/BRS.0b013e3181b61d10. Spine (Phila Pa 1976). 2009. PMID: 19881400
-
Incidence of subsequent vertebral fracture after kyphoplasty.Spine (Phila Pa 1976). 2004 Oct 15;29(20):2270-6; discussion 2277. doi: 10.1097/01.brs.0000142469.41565.2a. Spine (Phila Pa 1976). 2004. PMID: 15480139 Review.
Cited by
-
Biomechanical in vitro comparison of radiofrequency kyphoplasty and balloon kyphoplasty.Eur Spine J. 2017 Dec;26(12):3225-3234. doi: 10.1007/s00586-017-5035-5. Epub 2017 Apr 27. Eur Spine J. 2017. PMID: 28451858
-
Interdisciplinary approach to balloon kyphoplasty in the treatment of osteoporotic vertebral compression fractures.Eur Spine J. 2003 Oct;12 Suppl 2(Suppl 2):S163-7. doi: 10.1007/s00586-003-0591-2. Epub 2003 Sep 12. Eur Spine J. 2003. PMID: 13680311 Free PMC article. Review.
-
Percutaneous vertebroplasty as a treatment for osteoporotic vertebral compression fractures: a systematic review.Eur Spine J. 2006 Dec;15(12):1749-58. doi: 10.1007/s00586-006-0159-z. Epub 2006 Jul 6. Eur Spine J. 2006. PMID: 16823557
-
Effects of cement augmentation on the mechanical stability of multilevel spine after vertebral compression fracture.J Spine Surg. 2016 Jun;2(2):111-21. doi: 10.21037/jss.2016.06.05. J Spine Surg. 2016. PMID: 27683707 Free PMC article.
-
Load shift of the intervertebral disc after a vertebroplasty: a finite-element study.Eur Spine J. 2003 Aug;12(4):421-6. doi: 10.1007/s00586-002-0512-9. Epub 2003 Apr 1. Eur Spine J. 2003. PMID: 12687437 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical
Research Materials