Using digital versus analog ECG data in clinical trials
- PMID: 10688311
- DOI: 10.1016/s0022-0736(99)90057-2
Using digital versus analog ECG data in clinical trials
Abstract
In 54 patients, we measured 3 different electrocardiogram (ECG) parameters required in the Symphony 1 trial and compared various combinations of pairs of measurements from 4 different sets of analog data and 1 set of digital data. Particularly for measurements of the durations of R waves in lead aVF, we found poor intraobserver and interobserver reliability and poor agreement between analog and digital data. There was much better agreement in measurements of the amplitudes of S waves in lead V3. The poorer agreement involving durations appears to be due to difficulties in accurately identifying the offsets of portions of the QRS complex, especially in lead AVF. We conclude that, particularly in measurements of duration, digital and analog data are not equivalent.
Similar articles
-
Meet the challenge of high-pass filter and ST-segment requirements with a DC-coupled digital electrocardiogram amplifier.J Electrocardiol. 2009 Nov-Dec;42(6):574-9. doi: 10.1016/j.jelectrocard.2009.07.012. Epub 2009 Aug 22. J Electrocardiol. 2009. PMID: 19700169
-
Digital transmission of 12-lead electrocardiograms and duplex speech in the telephone bandwidth.J Telemed Telecare. 1996;2(1):42-9. doi: 10.1258/1357633961929150. J Telemed Telecare. 1996. PMID: 9375041 Clinical Trial.
-
Digitization of electrocardiograms by desktop optical scanner.J Electrocardiol. 1991 Oct;24(4):325-38. doi: 10.1016/0022-0736(91)90015-e. J Electrocardiol. 1991. PMID: 1744545
-
What is inside the electrocardiograph?J Electrocardiol. 2008 Jan-Feb;41(1):8-14. doi: 10.1016/j.jelectrocard.2007.08.059. J Electrocardiol. 2008. PMID: 18191652 Review.
-
Implications of methodological differences in digital electrocardiogram interval measurement.J Electrocardiol. 2006 Oct;39(4 Suppl):S152-6. doi: 10.1016/j.jelectrocard.2006.05.030. Epub 2006 Aug 21. J Electrocardiol. 2006. PMID: 16920141 Review.
Cited by
-
Magnitude, mechanism, and reproducibility of QT interval differences between superimposed global and individual lead ECG complexes.Ann Noninvasive Electrocardiol. 2007 Apr;12(2):145-52. doi: 10.1111/j.1542-474X.2007.00153.x. Ann Noninvasive Electrocardiol. 2007. PMID: 17593183 Free PMC article.
MeSH terms
LinkOut - more resources
Medical