Comparative Study
. 1999 Nov;52(11):807-11.
doi: 10.1136/jcp.52.11.807.

Clinical application of the Quantiplex HCV RNA 2.0 and Amplicor HCV Monitor assays for quantifying serum hepatitis C virus RNA

Affiliations
Comparative Study

Clinical application of the Quantiplex HCV RNA 2.0 and Amplicor HCV Monitor assays for quantifying serum hepatitis C virus RNA

M L Yu et al. J Clin Pathol. 1999 Nov.

Abstract

Aim: To compare the performance characteristics and clinical application of two different technologies for quantifying serum hepatitis C virus (HCV) RNA levels.

Methods: HCV RNA was quantified by Amplicor HCV Monitor assay (Amplicor) and Quantiplex HCV RNA 2.0 assay (bDNA-2) in 119 sera from 107 HCV infected patients.

Results: Both assays had similar sensitivity (79.4% for Amplicor; 86.0% for bDNA-2), acceptable coefficients of variation (5.3% in Amplicor; 2.6% in bDNA-2), and good linearity (r2 > or = 0.98). There was a positive correlation between quantification values of both methods (r = 0.683, p < 0.001). The Amplicor values were on an average 1.76 log lower than bDNA-2 results. Male subjects and HCV genotype 1b were significantly associated with higher viral load determined by Amplicor, but not with viral load measured by bDNA-2. In 70 chronic HCV infected patients treated with interferon alfa, mean (SD) pretreatment viral load in 27 complete responders (3.47 (0.84) logs for Amplicor, 5.63 (0.58) for bDNA-2) was significantly lower than in non-responders (4.43 (1.01) logs for Amplicor, 6.10 (0.67) logs for bDNA-2; p < 0.001). Cut off points of 3.9 logs for Amplicor and 5.8 logs for bDNA-2 were determined to be the best for predicting response to interferon alfa, giving acceptable sensitivity (70.4%, 74.1%), specificity (72.1%, 65.1%), and accuracy (71.4%, 68.6%), respectively.

Conclusions: Both the Amplicor and bDNA-2 assays are clinically useful methods for HCV RNA quantification and are reliable for predicting the outcome of treatment, despite differences in absolute quantification values and in the correlation between HCV genotypes and viral load.

PubMed Disclaimer

Similar articles

Cited by

References

    1. Radiology. 1983 Sep;148(3):839-43 - PubMed
    1. Lancet. 1998 Jan 10;351(9096):83-7 - PubMed
    1. Science. 1989 Apr 21;244(4902):362-4 - PubMed
    1. N Engl J Med. 1992 Dec 31;327(27):1899-905 - PubMed
    1. Gastroenterology. 1993 Mar;104(3):877-83 - PubMed

Publication types