Failure of an automated blood culture system to detect nonfermentative gram-negative bacteria
Abstract
During a 1-year study we observed that both aerobic and anaerobic blood culture bottles from patients were negative by the BacT/Alert system during a 7-day incubation period. However, upon subcultivation of negative bottles, growth of Pseudomonas aeruginosa was detectable. In an attempt to explain this observation, aerobic BacT/Alert Fan bottles were seeded with a defined inoculum (0.5 McFarland standard; 1 ml) of Escherichia coli, Klebsiella pneumoniae, Serratia marcescens, P. aeruginosa, Stenotrophomonas maltophilia, or Acinetobacter baumannii. Half of the inoculated bottles were loaded into the BacT/Alert system immediately, and the remainder were preincubated for 4, 8, 16, and 24 h at 36 degrees C. With preincubation all bottles seeded with the Enterobacteriaceae signaled positive during the next 1.5 h. Organisms in bottles seeded with the nonfermentative species P. aeruginosa and A. baumannii remained undetected by the BacT/Alert system for 7 days. S. maltophilia was detected if the preincubation time was equal or less than 8 h. Without preincubation all bottles seeded with the Enterobacteriaceae or nonfermentative species signaled positive. Since nonfermentative species seem to enter a state of bacteriostasis within the preincubation period, we reasoned that an unknown factor is consumed. Accordingly, a smaller inoculum should allow the detection of nonfermentative species, even after preincubation, and serial dilutions of P. aeruginosa were detected in preincubated bottles. In this case preincubated bottles signaled positive faster than bottles without preincubation. We conclude that all bottles from clinical settings should be subcultured prior to loading to avoid false negatives. An alternative may be preincubation at room temperature.
Figures

Comment in
-
Failure of an automated blood culture system to detect nonfermentative gram-negative bacteria.J Clin Microbiol. 2000 Jul;38(7):2803-4. doi: 10.1128/JCM.38.7.2803-2804.2000. J Clin Microbiol. 2000. PMID: 10979755 Free PMC article. No abstract available.
Similar articles
-
Comparative study of BacT/Alert FAN bottles and standard BacT/Alert bottles.Diagn Microbiol Infect Dis. 1998 Nov;32(3):159-63. doi: 10.1016/s0732-8893(98)00094-7. Diagn Microbiol Infect Dis. 1998. PMID: 9884831
-
Growth dynamics of Staphylococcus aureus, Escherichia coli, and Pseudomonas aeruginosa as a function of time to detection in BacT/alert 3D blood culture bottles with various preincubation conditions.Ann Lab Med. 2013 Nov;33(6):406-9. doi: 10.3343/alm.2013.33.6.406. Epub 2013 Oct 17. Ann Lab Med. 2013. PMID: 24205488 Free PMC article.
-
In-house validation of the BACTEC 9240 blood culture system for detection of bacterial contamination in platelet concentrates.Transfusion. 2005 Jul;45(7):1138-42. doi: 10.1111/j.1537-2995.2005.04343.x. Transfusion. 2005. PMID: 15987359
-
[Effect of delay of blood cultures on positive detection by automated blood culture system].Kansenshogaku Zasshi. 2004 Nov;78(11):959-66. doi: 10.11150/kansenshogakuzasshi1970.78.959. Kansenshogaku Zasshi. 2004. PMID: 15628528 Japanese.
-
Predominance of Enterobacteriaceae isolates in early positive anaerobic blood culture bottles in BacT/Alert system.J Clin Lab Anal. 2013 Mar;27(2):113-20. doi: 10.1002/jcla.21571. Epub 2013 Jan 24. J Clin Lab Anal. 2013. PMID: 23349061 Free PMC article.
Cited by
-
Failure of an automated blood culture system to detect nonfermentative gram-negative bacteria.J Clin Microbiol. 2000 Jul;38(7):2803-4. doi: 10.1128/JCM.38.7.2803-2804.2000. J Clin Microbiol. 2000. PMID: 10979755 Free PMC article. No abstract available.
-
Metagenomic evidence for a polymicrobial signature of sepsis.Microb Genom. 2021 Sep;7(9):000642. doi: 10.1099/mgen.0.000642. Microb Genom. 2021. PMID: 34477543 Free PMC article.
-
Impact of Pre-Analytical Time on the Recovery of Pathogens from Blood Cultures: Results from a Large Retrospective Survey.PLoS One. 2017 Jan 3;12(1):e0169466. doi: 10.1371/journal.pone.0169466. eCollection 2017. PLoS One. 2017. PMID: 28046040 Free PMC article.
-
Subculturing and Gram staining of blood cultures flagged negative by the BACTEC™ FX system: Optimizing the workflow for detection of Cryptococcus neoformans in clinical specimens.Front Microbiol. 2023 Mar 15;14:1113817. doi: 10.3389/fmicb.2023.1113817. eCollection 2023. Front Microbiol. 2023. PMID: 37007533 Free PMC article.
-
Apparent culture-negative prosthetic valve endocarditis caused by Peptostreptococcus magnus.J Clin Microbiol. 2000 Dec;38(12):4640-2. doi: 10.1128/JCM.38.12.4640-4642.2000. J Clin Microbiol. 2000. PMID: 11101611 Free PMC article.
References
-
- Pittet D, Li N, Woolson R F, Wenzel R P. Microbiological factors influencing the outcome of nosocomial bloodstream infections: a 6-year validated, population-based model. Clin Infect Dis. 1997;24:1068–1078. - PubMed
-
- Washington J A, II, Ilstrup D M. Blood cultures: issues and controversies. Rev Infect Dis. 1986;8:792–802. - PubMed
-
- Weinstein M P. Current blood culture methods and systems: clinical concepts, technology, and interpretation of results. Clin Infect Dis. 1996;23:40–46. - PubMed
MeSH terms
LinkOut - more resources
Full Text Sources
Medical