Failure of an automated blood culture system to detect nonfermentative gram-negative bacteria
- PMID: 10698992
- PMCID: PMC86332
- DOI: 10.1128/JCM.38.3.1036-1041.2000
Failure of an automated blood culture system to detect nonfermentative gram-negative bacteria
Abstract
During a 1-year study we observed that both aerobic and anaerobic blood culture bottles from patients were negative by the BacT/Alert system during a 7-day incubation period. However, upon subcultivation of negative bottles, growth of Pseudomonas aeruginosa was detectable. In an attempt to explain this observation, aerobic BacT/Alert Fan bottles were seeded with a defined inoculum (0.5 McFarland standard; 1 ml) of Escherichia coli, Klebsiella pneumoniae, Serratia marcescens, P. aeruginosa, Stenotrophomonas maltophilia, or Acinetobacter baumannii. Half of the inoculated bottles were loaded into the BacT/Alert system immediately, and the remainder were preincubated for 4, 8, 16, and 24 h at 36 degrees C. With preincubation all bottles seeded with the Enterobacteriaceae signaled positive during the next 1.5 h. Organisms in bottles seeded with the nonfermentative species P. aeruginosa and A. baumannii remained undetected by the BacT/Alert system for 7 days. S. maltophilia was detected if the preincubation time was equal or less than 8 h. Without preincubation all bottles seeded with the Enterobacteriaceae or nonfermentative species signaled positive. Since nonfermentative species seem to enter a state of bacteriostasis within the preincubation period, we reasoned that an unknown factor is consumed. Accordingly, a smaller inoculum should allow the detection of nonfermentative species, even after preincubation, and serial dilutions of P. aeruginosa were detected in preincubated bottles. In this case preincubated bottles signaled positive faster than bottles without preincubation. We conclude that all bottles from clinical settings should be subcultured prior to loading to avoid false negatives. An alternative may be preincubation at room temperature.
Figures
Comment in
-
Failure of an automated blood culture system to detect nonfermentative gram-negative bacteria.J Clin Microbiol. 2000 Jul;38(7):2803-4. doi: 10.1128/JCM.38.7.2803-2804.2000. J Clin Microbiol. 2000. PMID: 10979755 Free PMC article. No abstract available.
References
-
- Pittet D, Li N, Woolson R F, Wenzel R P. Microbiological factors influencing the outcome of nosocomial bloodstream infections: a 6-year validated, population-based model. Clin Infect Dis. 1997;24:1068–1078. - PubMed
-
- Washington J A, II, Ilstrup D M. Blood cultures: issues and controversies. Rev Infect Dis. 1986;8:792–802. - PubMed
-
- Weinstein M P. Current blood culture methods and systems: clinical concepts, technology, and interpretation of results. Clin Infect Dis. 1996;23:40–46. - PubMed
MeSH terms
LinkOut - more resources
Full Text Sources
Medical
