. 2000 Feb;45(2):511-28.
doi: 10.1088/0031-9155/45/2/317.

The measurement of proton stopping power using proton-cone-beam computed tomography

Affiliations

The measurement of proton stopping power using proton-cone-beam computed tomography

P Zygmanski et al. Phys Med Biol. 2000 Feb.

Abstract

A cone-beam computed tomography (CT) system utilizing a proton beam has been developed and tested. The cone beam is produced by scattering a 160 MeV proton beam with a modifier that results in a signal in the detector system, which decreases monotonically with depth in the medium. The detector system consists of a Gd2O2S:Tb intensifying screen viewed by a cooled CCD camera. The Feldkamp-Davis-Kress cone-beam reconstruction algorithm is applied to the projection data to obtain the CT voxel data representing proton stopping power. The system described is capable of reconstructing data over a 16 x 16 x 16 cm3 volume into 512 x 512 x 512 voxels. A spatial and contrast resolution phantom was scanned to determine the performance of the system. Spatial resolution is significantly degraded by multiple Coulomb scattering effects. Comparison of the reconstructed proton CT values with x-ray CT derived proton stopping powers shows that there may be some advantage to obtaining stopping powers directly with proton CT. The system described suggests a possible practical method of obtaining this measurement in vivo.

PubMed Disclaimer

Publication types

MeSH terms