Accuracy of oscillometric blood pressure measurement according to the relation between cuff size and upper-arm circumference in critically ill patients
- PMID: 10708169
- DOI: 10.1097/00003246-200002000-00014
Accuracy of oscillometric blood pressure measurement according to the relation between cuff size and upper-arm circumference in critically ill patients
Abstract
Objective: To evaluate the accuracy of oscillometric blood pressure measurement according to the relation between cuff size and upper-arm circumference in critically ill patients.
Design: Prospective data collection.
Setting: Emergency department in a 2,000-bed inner city hospital.
Patients: Thirty-eight patients categorized into three groups according to their upper-arm circumference (group I: 18-25 cm; group II: 25.1-33 cm; and group III: 33.1-47.5 cm) were enrolled in the study protocol.
Interventions: In each patient, all three cuff sizes (Hewlett-Packard Cuff 40401 B, C, and D) were used to perform an oscillometric blood pressure measurement at least within 3 mins until ten to 20 measurements for each cuff size were achieved. Invasive mean arterial blood pressure measurement was done by cannulation of the contralateral radial artery with direct transduction of the systemic arterial pressure waveform. The corresponding invasive blood pressure value was obtained at the end of each oscillometric measurement.
Measurement and main results: Overall, 1,494 pairs of simultaneous oscillometric and invasive blood pressure measurements were collected in 38 patients (group I, n = 5; group II, n = 23; and group III, n = 10) over a total time of 72.3 hrs. Mean arterial blood pressure ranged from 35 to 165 mm Hg. The overall discrepancy between oscillometric and invasive blood pressure measurement was -6.7+/-9.7 mm Hg (p<.0001), if the recommended cuff size according to the upper-arm circumference was used (539 measurements). Of all the blood pressure measurements, 26.4% (n = 395) had a discrepancy of > or =10 mm Hg and 34.2% (n = 512) exhibited a discrepancy of > or =20 mm Hg. No differences between invasive and noninvasive blood pressure measurements were noted in patients either with or without inotropic support (-6.6 + 7.2 vs. -8.6 + 6.8 mm Hg; not significant).
Conclusion: The oscillometric blood pressure measurement significantly underestimates arterial blood pressure and exhibits a high number of measurements out of the clinically acceptable range. The relation between cuff size and upper-arm circumference contributes substantially to the inaccuracy of the oscillometric blood pressure measurement. Therefore, oscillometric blood pressure measurement does not achieve adequate accuracy in critically ill patients.
Similar articles
-
Factors influencing the accuracy of oscillometric blood pressure measurement in critically ill patients.Crit Care Med. 2003 Mar;31(3):793-9. doi: 10.1097/01.CCM.0000053650.12025.1A. Crit Care Med. 2003. PMID: 12626986
-
Accuracy and reliability of noninvasive continuous finger blood pressure measurement in critically ill patients.Crit Care Med. 1996 Oct;24(10):1684-9. doi: 10.1097/00003246-199610000-00014. Crit Care Med. 1996. PMID: 8874306
-
Noninvasive monitoring of blood pressure in the critically ill: reliability according to the cuff site (arm, thigh, or ankle).Crit Care Med. 2012 Apr;40(4):1207-13. doi: 10.1097/CCM.0b013e31823dae42. Crit Care Med. 2012. PMID: 22425818
-
Method of Blood Pressure Measurement in Neonates and Infants: A Systematic Review and Analysis.J Pediatr. 2020 Jun;221:23-31.e5. doi: 10.1016/j.jpeds.2020.02.072. J Pediatr. 2020. PMID: 32446487
-
Cuff Under Pressure for Greater Accuracy.Curr Hypertens Rep. 2020 Sep 22;22(11):93. doi: 10.1007/s11906-020-01103-8. Curr Hypertens Rep. 2020. PMID: 32959103 Review.
Cited by
-
Functional hemodynamic monitoring.Crit Care. 2005;9(6):566-72. doi: 10.1186/cc3927. Epub 2005 Nov 22. Crit Care. 2005. PMID: 16356240 Free PMC article. Review.
-
A comparison of noninvasive blood pressure measurement on the wrist with invasive arterial blood pressure monitoring in patients undergoing bariatric surgery.Obes Surg. 2009 Jun;19(6):717-24. doi: 10.1007/s11695-008-9607-7. Epub 2008 Jul 10. Obes Surg. 2009. PMID: 18618207
-
Noninvasive techniques for blood pressure measurement are not a reliable alternative to direct measurement: a randomized crossover trial in ICU.ScientificWorldJournal. 2014 Jan 30;2014:353628. doi: 10.1155/2014/353628. eCollection 2014. ScientificWorldJournal. 2014. PMID: 24616624 Free PMC article. Clinical Trial.
-
Arterial Pressure Monitoring in Pediatric Patients Undergoing Cardiac Surgery: An Observational Study Comparing Invasive and Non-invasive Measurements.Pediatr Cardiol. 2019 Aug;40(6):1231-1237. doi: 10.1007/s00246-019-02137-9. Epub 2019 Jun 20. Pediatr Cardiol. 2019. PMID: 31222376
-
Noninvasive continuous hemodynamic monitoring.J Clin Monit Comput. 2012 Aug;26(4):267-78. doi: 10.1007/s10877-012-9375-8. Epub 2012 Jun 14. J Clin Monit Comput. 2012. PMID: 22695821 Free PMC article. Review.
MeSH terms
LinkOut - more resources
Full Text Sources
Medical