Anterior capsule opacification: a histopathological study comparing different IOL styles

Affiliations
Comparative Study

Anterior capsule opacification: a histopathological study comparing different IOL styles

L Werner et al. Ophthalmology. 2000 Mar.

Abstract

Objective: To compare the degree of anterior capsule opacification (ACO) in human eyes obtained post-mortem containing various rigid and foldable posterior chamber intraocular lens (PC-IOL) designs.

Design: Comparative autopsy tissue study with clinicopathologic correlations.

Materials: Four hundred sixty human globes containing the following PC-IOL styles were analyzed: (1) one-piece polymethylmethacrylate (PMMA) optic-PMMA haptic (n = 50), (2) one-piece silicone-plate IOL, large hole (n = 40), (3) one-piece silicone-plate IOL, small hole (n = 67), (4) three-piece PMMA optic-PMMA/Prolene haptic (n = 51), (5) three-piece acrylic optic-PMMA haptic (n = 96), (6) three-piece silicone optic-PMMA haptic (n = 24), (7) three-piece silicone optic-polyimide haptic (n = 40), and (8) three-piece silicone optic-prolene haptic (n = 92).

Testing: The globes were sectioned in the equatorial plane for gross examination and then processed through paraffin; sectioned, and stained with hematoxylin-eosin, periodic acid-Schiff, and Masson's trichrome stains; and examined by light microscopy.

Main outcome measures: Anterior capsule opacification was scored in each eye by grading the histologic sections from 0 to III, according to the amount (thickness) of proliferative tissue and cells measured in sagittal sections on the inner surface of the anterior capsule at the capsulorhexis margin.

Results: The difference among the eight groups was significant (P < 0.0001). Mean ACO scores were highest with the large and small hole one-piece silicone-plate lenses (1.77 +/- 0.86 and 1.28 +/- 0.77, respectively). The lowest mean score was observed in the group of three-piece acrylic optic-PMMA haptics lenses (0.51 +/- 0.52).

Conclusions: Our results confirm previous clinical observations that the rate of ACO is relatively high with plate-haptic silicone IOLs. The lowest rate was noted with the three-piece acrylic optic-PMMA haptic IOL. The IOL design and IOL material are significant factors in the development of ACO.