Effects of luminance and resolution on observer performance with chest radiographs
- PMID: 10751483
- DOI: 10.1148/radiology.215.1.r00ap34169
Effects of luminance and resolution on observer performance with chest radiographs
Abstract
Purpose: To examine the combined effects of image resolution and display luminance on observer performance for detection of abnormalities depicted on posteroanterior chest radiographs.
Materials and methods: A total of 529 radiographs were displayed on a specially constructed view box at three luminance levels (770, 260, and 85 cd/m(2)) and three resolutions (100-microm, 200-microm, and 400-microm pixels). Each image was reviewed nine times by six radiologists who participated in this study. The abnormalities included nodule, pneumothorax, interstitial disease, alveolar infiltrates, and rib fracture. Negative (normal) radiographs were also included.
Results: Receiver operating characteristic curves indicated that the effect of image luminance was greater than that of resolution. The detection of pneumothorax, interstitial disease, and rib fracture showed statistically significant differences (P <. 05) due to luminance. The detection of pneumothorax was the only abnormality with a statistically significant difference due to resolution. There was no evidence that luminance was related to image resolution for any of the abnormalities.
Conclusion: At a resolution of 400-microm pixels or higher across the field of view and a luminance of 260 cd/m(2) or more, primary diagnosis with posteroanterior chest radiographs is not likely to be affected by the quality of display.