American College of Cardiology/European Society of Cardiology International Study of Angiographic Data Compression Phase III: measurement of image quality differences at varying levels of data compression
- PMID: 10758989
- DOI: 10.1016/s0735-1097(99)00655-5
American College of Cardiology/European Society of Cardiology International Study of Angiographic Data Compression Phase III: measurement of image quality differences at varying levels of data compression
Abstract
Objectives: We sought to investigate up to which level of Joint Photographic Experts Group JPEG) data compression the perceived image quality and the detection of diagnostic features remain equivalent to the quality and detectability found in uncompressed coronary angiograms.
Background: Digital coronary angiograms represent an enormous amount of data and therefore require costly computerized communication and archiving systems. Earlier studies on the viability of medical image compression were not fully conclusive.
Methods: Twenty-one raters evaluated sets of 91 cine runs. Uncompressed and compressed versions of the images were presented side by side on one monitor, and image quality differences were assessed on a scale featuring six scores. In addition, the raters had to detect pre-defined clinical features. Compression ratios (CR) were 6:1, 10:1 and 16:1. Statistical evaluation was based on descriptive statistics and on the equivalence t-test.
Results: At the lowest CR (CR 6:1), there was already a small (15%) increase in assigning the aesthetic quality score indicating "quality difference is barely discernible-the images are equivalent." At CR 10:1 and CR 16:1, close to 10% and 55%, respectively, of the compressed images were rated to be "clearly degraded, but still adequate for clinical use" or worse. Concerning diagnostic features, at CR 10:1 and CR 16:1 the error rate was 9.6% and 13.1%, respectively, compared with 9% for the baseline error rate in uncompressed images.
Conclusions: Compression at CR 6:1 provides equivalence with the original cine runs. If CR 16:1 were used, one would have to tolerate a significant increase in the diagnostic error rate over the baseline error rate. At CR 10:1, intermediate results were obtained.
Similar articles
-
American College of Cardiology/ European Society of Cardiology international study of angiographic data compression phase III. Measurement Of image quality differences at varying levels of data compression.Eur Heart J. 2000 Apr;21(8):687-96. doi: 10.1053/euhj.1999.2102. Eur Heart J. 2000. PMID: 10731407 Clinical Trial.
-
American College of Cardiology/European Society of Cardiolgoy International Study of Angiographic Data Compression Phase II: the effects of varying JPEG data compression levels on the quantitative assessment of the degree of stenosis in digital coronary angiography. Joint Photographic Experts Group.J Am Coll Cardiol. 2000 Apr;35(5):1380-7. doi: 10.1016/s0735-1097(99)00611-7. J Am Coll Cardiol. 2000. PMID: 10758988 Clinical Trial.
-
American College of Cardiology/European Society of Cardiology International Study of Angiographic Data Compression Phase I: The effect of lossy data compression on recognition of diagnostic features in digital coronary angiography.J Am Coll Cardiol. 2000 Apr;35(5):1370-9. doi: 10.1016/s0735-1097(99)00610-5. J Am Coll Cardiol. 2000. PMID: 10758987 Clinical Trial.
-
European Association of Cardiovascular Imaging/Cardiovascular Imaging Department of the Brazilian Society of Cardiology recommendations for the use of cardiac imaging to assess and follow patients after heart transplantation.Eur Heart J Cardiovasc Imaging. 2015 Sep;16(9):919-48. doi: 10.1093/ehjci/jev139. Epub 2015 Jul 2. Eur Heart J Cardiovasc Imaging. 2015. PMID: 26139361 Review.
-
Quality criteria for cardiac images: an update.Radiat Prot Dosimetry. 2008;129(1-3):87-90. doi: 10.1093/rpd/ncn034. Epub 2008 Feb 18. Radiat Prot Dosimetry. 2008. PMID: 18283064 Review.
Cited by
-
Influence of image metrics when assessing image quality from a test object in cardiac X-ray systems: Part II.J Digit Imaging. 2012 Aug;25(4):537-41. doi: 10.1007/s10278-011-9448-0. J Digit Imaging. 2012. PMID: 22223157 Free PMC article.
-
Synchronized, interactive teleconferencing with digital cardiac images.J Digit Imaging. 2006 Mar;19(1):85-91. doi: 10.1007/s10278-005-8147-0. J Digit Imaging. 2006. PMID: 16249837 Free PMC article.
-
Comparison of image compression viability for lossy and lossless JPEG and Wavelet data reduction in coronary angiography.Int J Cardiovasc Imaging. 2001 Feb;17(1):1-12. doi: 10.1023/a:1010644318298. Int J Cardiovasc Imaging. 2001. PMID: 11495503
-
Influence of image metrics when assessing image quality from a test object in cardiac X-ray systems.J Digit Imaging. 2011 Apr;24(2):331-8. doi: 10.1007/s10278-009-9268-7. J Digit Imaging. 2011. PMID: 20127268 Free PMC article.