A new algorithm for the quantitation of myocardial perfusion SPECT. I: technical principles and reproducibility
- PMID: 10768574
A new algorithm for the quantitation of myocardial perfusion SPECT. I: technical principles and reproducibility
Abstract
We have developed a new, completely automatic 3-dimensional software approach to quantitative perfusion SPECT. The main features of the software are myocardial sampling based on an ellipsoid model; use of the entire count profile between the endocardial and epicardial surfaces; independence of the algorithm from myocardial shape, size, and orientation and establishment of a standard 3-dimensional point-to-point correspondence among all sampled myocardial regions; automatic generation of quantitative measurements and 5-point semiquantitative scores for each of 20 myocardial segments and automatic derivation of summed perfusion scores; and automatic generation of normal limits for any given patient population on the basis of data fractionally normalized to minimize hot spot artifacts.
Methods: The new algorithm was tested on the tomographic images of 420 patients studied with a rest 201TI (111-167 MBq, 35 s/projection)-stress 99mTc-sestamibi (925-1480 MBq, 25 s/projection) separate dual-isotope protocol on a single-detector camera, a dual-detector 90 degrees camera, and a triple-detector camera.
Results: The algorithm was successful in 397 of 420 patients (94.5%) and 816 of 840 image datasets (97.1%), with a statistically significant difference between the success rates of the 201TI images (399/ 420, or 95.0%) and the 99mTc images (417/420, or 99.3%; P < 0.001). Algorithm failure was caused by extracardiac uptake (10/24, or 41.7%) or inaccurate identification of the valve plane because of low count statistics (14/24, or 58.3%) and was obviated by simply limiting the image volume in which the software operates. Reproducibility of measurements of summed perfusion scores (r = 0.999 and 1 for stress and rest, respectively), global defect extent (r = 0.999 and 1 for stress and rest, respectively), and segmental perfusion scores (exact agreement = 99.9%, kappa = 0.998 for stress and 0.997 for rest) was extremely high.
Conclusion: Automatic 3-dimensional quantitation of perfusion from 201Tl and 99mTc-sestamibi images is feasible and reproducible. The described software, because it is based on the same sampling scheme used for gated SPECT analysis, ensures intrinsically perfect registration of quantitative perfusion with quantitative regional wall motion and thickening information, if gated SPECT is used.
Comment in
-
New algorithm for quantification of myocardial perfusion SPECT.J Nucl Med. 2001 Feb;42(2):391-2. J Nucl Med. 2001. PMID: 11216542 No abstract available.
Similar articles
-
A new algorithm for the quantitation of myocardial perfusion SPECT. II: validation and diagnostic yield.J Nucl Med. 2000 Apr;41(4):720-7. J Nucl Med. 2000. PMID: 10768575
-
Automatic quantitation of regional myocardial wall motion and thickening from gated technetium-99m sestamibi myocardial perfusion single-photon emission computed tomography.J Am Coll Cardiol. 1997 Nov 1;30(5):1360-7. doi: 10.1016/s0735-1097(97)00276-3. J Am Coll Cardiol. 1997. PMID: 9350940
-
Feasibility of simultaneous stress 99mTc-sestamibi/rest 201Tl dual-isotope myocardial perfusion SPECT in the detection of coronary artery disease.J Nucl Med. 1999 Jun;40(6):895-903. J Nucl Med. 1999. PMID: 10452303
-
An automatic approach to the analysis, quantitation and review of perfusion and function from myocardial perfusion SPECT images.Int J Card Imaging. 1997 Aug;13(4):337-46. doi: 10.1023/a:1005815206195. Int J Card Imaging. 1997. PMID: 9306148 Review.
-
Technetium 99m sestamibi in the assessment of chronic coronary artery disease.Semin Nucl Med. 1991 Jul;21(3):190-212. doi: 10.1016/s0001-2998(05)80040-6. Semin Nucl Med. 1991. PMID: 1835137 Review.
Cited by
-
Transient ischemic dilation in SPECT myocardial perfusion imaging for prediction of severe coronary artery disease in diabetic patients.J Nucl Cardiol. 2013 Feb;20(1):45-52. doi: 10.1007/s12350-012-9642-6. Epub 2012 Oct 23. J Nucl Cardiol. 2013. PMID: 23090352
-
Impact of imaging protocol on left ventricular ejection fraction using gated-SPECT myocardial perfusion imaging.J Nucl Cardiol. 2017 Aug;24(4):1292-1301. doi: 10.1007/s12350-016-0483-6. Epub 2016 Apr 6. J Nucl Cardiol. 2017. PMID: 27052809
-
Clinical Validation of Japanese Normal Myocardial Perfusion Imaging Databases Using Semi-conductor Gamma Camera (D-SPECT): Japanese Society of Nuclear Cardiology Working Group Reports.Ann Nucl Cardiol. 2022;8(1):36-41. doi: 10.17996/anc.22-00163. Epub 2022 Aug 31. Ann Nucl Cardiol. 2022. PMID: 36540179 Free PMC article.
-
The impact of the excimer laser on myocardial salvage in ST-elevation acute myocardial infarction via nuclear scintigraphy.Int J Cardiovasc Imaging. 2020 Jan;36(1):161-170. doi: 10.1007/s10554-019-01690-x. Epub 2019 Aug 26. Int J Cardiovasc Imaging. 2020. PMID: 31451993
-
Degradation of myocardial perfusion SPECT images caused by contaminants in thallous (201Tl) chloride.Eur J Nucl Med Mol Imaging. 2008 May;35(5):922-32. doi: 10.1007/s00259-008-0719-z. Epub 2008 Jan 25. Eur J Nucl Med Mol Imaging. 2008. PMID: 18219482
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources