Comparison of the MB/BacT and BACTEC MGIT 960 system for recovery of mycobacteria from clinical specimens

Affiliations
Comparative Study

Comparison of the MB/BacT and BACTEC MGIT 960 system for recovery of mycobacteria from clinical specimens

J J Yan et al. Diagn Microbiol Infect Dis. 2000 May.

Abstract

A total of 543 specimens were cultured in parallel with the MB/BacT and BACTEC MGIT 960 systems and on the conventional solid media. Mycobacteria were identified from 95 (17.5%) specimens, including 63 (66.3%) Mycobacterium tuberculosis and 32 (33.7%) nontuberculous mycobacteria. The recovery rates for the MB/BacT, MGIT 960, and solid media were 91.6, 87.4, and 54.7%, respectively, for all mycobacteria; the recovery rates were 93.6, 88.9, and 63.4%, respectively, for M. tuberculosis complex alone, and 87.5, 84.4, and 37.5%, respectively, for all nontuberculous mycobacteria. The mean times to detection of all mycobacteria by individual systems were 13. 9, 8.7, 31.7 days for the MB/BacT, MGIT 960 and solid media, respectively, 13.9, 9.3, 32.9 days for M. tuberculosis alone, and 14. 1, 8.1, 27.2 days for all nontuberculous mycobacteria. The contamination rates of the MB/BacT and MGIT 960 were 10.2 and 5.4%, respectively. With regard to detection times and recovery rates, both automated systems are superior to the conventional media (all p < 0.005). As compared to the MB/BacT, the MGIT 960 detected mycobacterial growth more rapidly (p < 0.001), and had a lower contamination rate (p = 0.003); however, there was no statistically significant difference in recovery rates between these two systems. These results indicate that both MGIT 960 and MB/BacT systems are rapid, sensitive, and efficient methods for the recovery of mycobacteria from clinical specimens.