Multifrequency method for dielectric monitoring of cold-preserved organs
Abstract
To answer a growing need for non-invasive monitoring of biological organs, we have developed an automated system capable of repeated dielectric measurements over the frequency range 10 kHz-100 MHz. Further, we propose a novel method of data analysis that may convert the acquired, individual dispersion curves into a diagram of the time course of specific phenomenological parameters, such as the characteristic frequency. By using this new procedure, unattended, long-term monitoring of temporal changes in the dielectric behaviour of excised liver lobes stored at 4 degrees C was successfully realized. The 'multifrequency' method presented here was definitely superior to the conventional 'fixed-frequency' method in providing reliable results.