Total hip arthroplasty with use of the Metasul metal-on-metal articulation. Four to seven-year results
- PMID: 10859098
- DOI: 10.2106/00004623-200006000-00005
Total hip arthroplasty with use of the Metasul metal-on-metal articulation. Four to seven-year results
Abstract
Background: Total hip replacements with a metal-on-metal articulation were commonly used until the mid-1970s; most were then abandoned in favor of hip replacement with a metal-on-polyethylene articulation. The reason for this change was primarily early cup loosening, which was more prevalent with these metal-on-metal designs than it was with metal-on-polyethylene designs. In the late 1980s, a metal-on-metal design with improved clearance (adequate space between the femoral head and the acetabular articulation surface to allow fluid film lubrication and clearance of any debris from within this joint), metal hardness, and reproducible surfaces was introduced by Sulzer Orthopedics in Switzerland. Orthopaedic surgeons were interested in this Metasul articulation because the contribution of polyethylene wear particles to the failure of total hip replacements had become evident. This study was undertaken to review the clinical performance of this implant and to determine if early acetabular loosening or revision and wear and osteolysis were prevalent.
Methods: Between 1991 and 1994, seventy patients (seventy hips) had a total hip replacement with the Metasul metal-on-metal articulation and a cemented Weber cup. Nine patients died less than four years after the replacement; none of these deaths were related to the operation. Five patients were not available for radiographic evaluation, but they were contacted and it was known that the hip was not painful and had not been revised. Fifty-six patients (fifty-six hips) had complete clinical and radiographic data four to 6.8 years after the operation, and they made up the study group. The patients were evaluated with use of the Harris hip score, a patient-self-assessment form, and radiographs.
Results: At an average of 5.2 years (range, four to 6.8 years) after the operation, the average total Harris hip score for the fifty-three patients who did not have a revision was 89.6 points (range, 62 to 100 points). The average Harris pain score was 41.0 points (range, 30 to 44 points), and the average Harris limp score was 9.4 points (range, 5 to 11 points). One patient had revision of a loose cup, but there were no other loose acetabular components in the series. Two patients had revision of the acetabular component because of dislocation. No patient had a loose or revised femoral component. Therefore, the mechanical failure rate was one (2 percent) of fifty-six patients. Thirty-six of forty-seven patients who completed the patient-self-assessment form rated their result as excellent; seven, as very good; two, as good; one, as fair; and one, as poor. Wear could not be measured on radiographs because of the metal-on-metal articulation. No hip had radiographic evidence of acetabular osteolysis and two hips had calcar resorption, but there was no other radiographic evidence of focal osteolysis.
Conclusions: Our four to seven-year experience with this articulation surface indicates that the clinical results are similar to those of total hip replacements with a metal-on-polyethylene articulation. We believe that the Metasul articulation may have a role in reducing the wear that occurs with total hip replacement. The Metasul articulation appears to be particularly indicated for more active patients. A historical comparison with the reports in the literature of which we are aware indicated that the hips in our study had a lower rate of acetabular revision and loosening than did those with previous metal-on-metal designs and that they had no more acetabular loosening or osteolysis than did those with metal-on-polyethylene articulations followed for an average of five years.
Comment in
-
Prosthetic design and outcome in total hip arthroplasty.J Bone Joint Surg Am. 2001 May;83(5):783-5. doi: 10.2106/00004623-200105000-00024. J Bone Joint Surg Am. 2001. PMID: 11379752 No abstract available.
Similar articles
-
The Harris-Galante porous-coated acetabular component with screw fixation. An average ten-year follow-up study.J Bone Joint Surg Am. 1999 Jan;81(1):66-73. doi: 10.2106/00004623-199901000-00010. J Bone Joint Surg Am. 1999. PMID: 9973056
-
High rate of revision and a high incidence of radiolucent lines around Metasul metal-on-metal total hip replacements: results from a randomised controlled trial of three bearings after seven years.Bone Joint J. 2013 Jul;95-B(7):881-6. doi: 10.1302/0301-620X.95B7.31067. Bone Joint J. 2013. PMID: 23814237 Clinical Trial.
-
Second-generation cementless total hip arthroplasty. Eight to eleven-year results.J Bone Joint Surg Am. 2001 Nov;83(11):1666-73. doi: 10.2106/00004623-200111000-00008. J Bone Joint Surg Am. 2001. PMID: 11701789
-
The argument for the use of Metasul as an articulation surface in total hip replacement.Clin Orthop Relat Res. 2004 Dec;(429):80-5. doi: 10.1097/01.blo.0000150343.66755.79. Clin Orthop Relat Res. 2004. PMID: 15577470 Review.
-
[Total hip arthroplasty revision surgery with migration of acetabular component to the subperitoneal space. Case report and literature review].Acta Ortop Mex. 2025 Mar-Apr;39(2):108-116. Acta Ortop Mex. 2025. PMID: 40335128 Review. Spanish.
Cited by
-
2008 John Charnley award: metal ion levels after metal-on-metal total hip arthroplasty: a randomized trial.Clin Orthop Relat Res. 2009 Jan;467(1):101-11. doi: 10.1007/s11999-008-0540-9. Epub 2008 Oct 15. Clin Orthop Relat Res. 2009. PMID: 18855089 Free PMC article. Clinical Trial.
-
Small-head metal on metal total hip arthroplasty is associated with a high rate of complication and reoperation at mid-term follow-up.SAGE Open Med. 2021 May 16;9:20503121211014707. doi: 10.1177/20503121211014707. eCollection 2021. SAGE Open Med. 2021. PMID: 34035911 Free PMC article.
-
Are Serum Metal Ion Levels a Concern at Mid-term Followup of Revision Knee Arthroplasty With a Metal-on-metal Hinge Design?Clin Orthop Relat Res. 2019 Sep;477(9):2007-2014. doi: 10.1097/CORR.0000000000000638. Clin Orthop Relat Res. 2019. PMID: 30811355 Free PMC article.
-
Cementless large-head metal-on-metal total hip arthroplasty in patients younger than 60 years--a multicenter early result.Kaohsiung J Med Sci. 2012 Jan;28(1):30-7. doi: 10.1016/j.kjms.2011.06.021. Epub 2011 Sep 15. Kaohsiung J Med Sci. 2012. PMID: 22226059 Free PMC article.
-
Metal-on-Metal Total Hip Resurfacing Arthroplasty: An Evidence-Based Analysis.Ont Health Technol Assess Ser. 2006;6(4):1-57. Epub 2006 Feb 1. Ont Health Technol Assess Ser. 2006. PMID: 23074495 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical