Microbiologic purity of dialysate: rationale and technical aspects
- PMID: 10859423
- DOI: 10.1159/000014419
Microbiologic purity of dialysate: rationale and technical aspects
Abstract
Dialysate purity has become a major concern in hemodialysis since it has been shown that microbial-derived products were stimulating the production and the release of proinflammatory cytokines in hemodialysis patients. This chronic microinflammatory state induced by hemodialysis has been putatively implicated in the development of dialysis-related pathology. In order to prevent risk related to these offenders and to reduce patient/dialysis interaction, it appears highly desirable to use ultrapure dialysis fluid aiming at sterility and apyrogenicity on a regular basis. Ultrapure dialysate results from a complex chain of production where purity grade relies on the weaker link of this chain. Technical aspects and pitfalls in the production of ultrapure dialysate are summarized in this paper. Production of ultrapure dialysate may be achieved on a routine basis, provided adequate components are used, and hygienic handling is regularly ensured. It includes the use of ultrapure water, clean and or sterile electrolytic concentrates (liquid or powder), implementation of ultrafilters on hemodialysis machines, microbiologic monitoring and hygienic handling of the chain with frequent disinfection. Safety and reliability of ultrapure dialysate production relies on a continuous quality assurance process, where results are coupled to corrective action in a feedback loop process.
Copyright 2000 S. Karger AG, Basel
Similar articles
-
Monitoring the microbial purity of the treated water and dialysate.Saudi J Kidney Dis Transpl. 2001 Jul-Sep;12(3):325-6. Saudi J Kidney Dis Transpl. 2001. PMID: 18209379
-
Microbiological purity of dialysate for on-line substitution fluid preparation.Nephrol Dial Transplant. 2000;15 Suppl 2:21-30. doi: 10.1093/ndt/15.suppl_1.21. Nephrol Dial Transplant. 2000. PMID: 11051034 Review.
-
[The concept of ultrapure solution online production].G Ital Nefrol. 2008 May-Jun;25 Suppl 41:S28-32, discussion S33-44. G Ital Nefrol. 2008. PMID: 18473318 Italian.
-
Ultrapure dialysate for home hemodialysis?Adv Chronic Kidney Dis. 2007 Jul;14(3):256-62. doi: 10.1053/j.ackd.2007.03.007. Adv Chronic Kidney Dis. 2007. PMID: 17603979 Review.
-
Optimal therapeutic conditions for online hemodiafiltration.Contrib Nephrol. 2011;168:28-38. doi: 10.1159/000321742. Epub 2010 Oct 7. Contrib Nephrol. 2011. PMID: 20938123
Cited by
-
A randomized cross-over study comparing the performance of HD integra™ central concentrate system versus pre-produced concentrate in hemodialysis.BMC Nephrol. 2017 Apr 3;18(1):123. doi: 10.1186/s12882-017-0537-2. BMC Nephrol. 2017. PMID: 28372556 Free PMC article. Clinical Trial.
-
Frequent hemodialysis with NxStage system in pediatric patients receiving maintenance hemodialysis.Pediatr Nephrol. 2008 Jan;23(1):129-35. doi: 10.1007/s00467-007-0649-1. Epub 2007 Nov 8. Pediatr Nephrol. 2008. PMID: 17990005
-
Pseudo-outbreak of haemodiafiltration dialysis fluid contamination: results of a detailed epidemiologic investigation.Infect Prev Pract. 2023 Jun 14;5(3):100292. doi: 10.1016/j.infpip.2023.100292. eCollection 2023 Sep. Infect Prev Pract. 2023. PMID: 37692534 Free PMC article.
-
Ultrapure dialysis fluid: a new standard for contemporary hemodialysis.Nephrourol Mon. 2012 Summer;4(3):519-23. doi: 10.5812/numonthly.3060. Epub 2012 Jun 20. Nephrourol Mon. 2012. PMID: 23573478 Free PMC article. No abstract available.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources