. 2000 Jul;69(1):82-6.
doi: 10.1136/jnnp.69.1.82.

Camino intracranial pressure monitor: prospective study of accuracy and complications

Affiliations

Camino intracranial pressure monitor: prospective study of accuracy and complications

R M Martínez-Mañas et al. J Neurol Neurosurg Psychiatry. 2000 Jul.

Abstract

Objectives: The fibreoptic device is a type of intracranial pressure monitor which seems to offer certain advantages over conventional monitoring systems. This study was undertaken to analyse the accuracy, drift characteristics, and complications of the Camino fibreoptic device.

Methods: One hundred and eight Camino intracranial pressure (ICP) devices, in their three modalities, were implanted during 1997. The most frequent indication for monitoring was severe head injury due to road traffic accidents.

Results: Sixty eight probe tips were cultured; 13.2% of the cases had a positive culture without clinical signs of infection, and 2.9% had a positive culture with clinical signs of ventriculitis. The most common isolated pathogen was Staphylococcus epidermidis. All patients were under cephalosporin prophylaxis during monitoring. Haemorrhage rate in patients without coagulation disorders was 2.1% and 15.3% in patients with coagulation abnormalities. Drift characteristics were studied in 56 cases; there was no drifting from the values expected according to the manufacturer's specifications in 34 probes. There was no relation between direction of the drift and duration of placement, nor between drift and time.

Conclusions: Although the complication and drift rates were similar to those reported elsewhere, there was no correlation between the direction of the drift and long term monitoring despite the fact that some published papers refer to overestimation of values with time with this type of device.

PubMed Disclaimer

Comment in

  • Statistics and analysis of the Camino ICP monitor.
    Gregson BA, Banister K, Chambers IR. Gregson BA, et al. J Neurol Neurosurg Psychiatry. 2001 Jan;70(1):138. doi: 10.1136/jnnp.70.1.138a. J Neurol Neurosurg Psychiatry. 2001. PMID: 11200140 Free PMC article. No abstract available.

References

    1. Lancet. 1993 Jan 16;341(8838):157-8 - PubMed
    1. J Hepatol. 1993 Jun;18(2):205-9 - PubMed
    1. Clin Pediatr (Phila). 1972 Apr;11(4):201-4 - PubMed
    1. J Neurosurg. 1976 May;44(5):567-70 - PubMed
    1. Clin Neurosurg. 1976;23:494-519 - PubMed